MN220-1 B 63
Portable Oxygen Concentrator
Figure 14.4: Tightening Center Bolt
COMPRESSOR ASSEMBLY INSTALLATION
To install the new compressor assembly, perform the compressor removal procedure in reverse
order.
After replacing the unit’s compressor, you must reset the hour meter. For
a FreeStyle, please reference the HOUR METER FUNCTIONALITY section of
this manual. If replacing the compressor on a FreeStyle 5, you must refer
to the instructions for reading and recording hours in the Hour Meter Kit -
Part Number KI406-1.
PRODUCT MANIFOLD REPLACEMENT
1.
Remove the PSA system per the PSA REMOVAL section in this manual.
2.
Disconnect the tubing assembly from the barb(s) on the product manifold, mixing tank, and
one end of the purge tube.