CAEN
Electronic Instrumentation
UM2619 – N1419 User Manual
33
Meaning of Board Alarm bits
Bit
Function
Bit 0 →
CH0
1 : Ch0 in Alarm status
Bit 1 →
CH1
1 : Ch1 in Alarm status
Bit 2 →
CH2
1 : Ch2 in Alarm status
Bit 3 → CH3
1 : Ch3 in Alarm status
Bit 4 → PWFAIL
1 : Board in POWER FAIL
Bit 5 → OVP
1 : Board in OVER POWER
Bit 6 →
HVCKFAIL
1 : Internal HV Clock FAIL (
≠
200±10kHz)
SET commands related to the Channels
The following table contains the strings to be used to handle set commands related to the Channels.
The '
X
' in the Field 'Channel' can be set to the '
0..(N-1)'
values.
7
When '
X=N
' the command is issued to all Channels.
String
Function (Units)
$BD:xx,CMD:SET,CH:X,PAR:VSET,VAL:XXXX.X
Set VSET value
$BD:xx,CMD:SET,CH:X,PAR:ISET,VAL:XXXX.XX
Set ISET value
$BD:xx,CMD:SET,CH:X,PAR:MAXV,VAL:XXXX
Set MAXVSET value
$BD:xx,CMD:SET,CH:X,PAR:RUP,VAL:XXX
Set RAMP UP value
$BD:xx,CMD:SET,CH:X,PAR:RDW,VAL:XXX
Set RAMP DOWN value
$BD:xx,CMD:SET,CH:X,PAR:TRIP,VAL:XXXX.X
Set TRIP time value
$BD:xx,CMD:SET,CH:X,PAR:PDWN,VAL:RAMP/KILL
Set POWER DOWN mode value
$BD:xx,CMD:SET,CH:X,PAR:IMRANGE,VAL:HIGH/LOW
Set IMON RANGE value
8
$BD:xx,CMD:SET,CH:X,PAR:ON
Set Ch ON
$BD:xx,CMD:SET,CH:X,PAR:OFF
Set Ch OFF
SET commands related to the module
String
Function (Units)
$BD:xx,CMD:SET,PAR:BDILKM,VAL:OPEN/CLOSED
Set Interlock Mode
$BD:xx,CMD:SET,PAR:BDCLR
Clear alarm signal
7
N
is the number of channels
8
parameter 'IMRANGE' can be changed only on modules featuring IMON zoom (optional)