background image

CAEN

 

 

Electronic Instrumentation

 

 

803x Series 8-16 Ch. Programmable HV Power Supplies

 

12 

Technical specifications table

1

 

Series 

8031 

8032 

8033 

8034 

8034H 

Polarity 

Positive / Negative / Mixed

2

 depending on purchased version; common ground 

Output connector 

BNC 

SHV 

Output Voltage 

0÷100 V 

0÷500 V 

0÷4 kV 

0÷6 kV 

Max. Output Current 

10 mA 

10mA 

3 mA 

1 mA 

20 µA 

Max. Output Power 

1 W 

5 W 

6 W 

0.12 W 

Voltage Set Resolution 

500µV

 

2mV

 

10mV

 

20mV

 

Voltage Monitor Resolution 

500µV

 

2mV

 

10mV

 

20mV

 

ISet Resolution 

50nA

 

10nA

 

5nA

 

100pA

 

IMon 
Resolution 

High range 

50nA

 

10nA

 

5nA

 

100pA

 

Low range 

5nA

 

1nA

 

500pA

 

10pA

 

Ramp Up/Down 

1÷50 V/s, 1 V/s step 

1÷100 V/s, 1 V/s step 

1÷500 V/s, 1 V/s step 

VMAX hardware 

0÷100 V 

0÷500 V 

0÷4 kV 

0÷6 kV 

VMAX hardware resolution 

1 V 

2V 

IMAX hardware 

0÷10 mA 

0÷3 mA 

0÷1 mA 

0÷20 µA 

IMAX hardware resolution 

10µA 

1µA 

20nA 

Alarm output 

Open collector, 100 mA maximum sink current 

Interlock input 

LOW: <1V; current~5mA; HIGH: 4÷6 V 

Trip 

Max. time an "overcurrent" can last (seconds). A channel in "overcurrent" works as a current generator; 
output voltage varies to keep the output current lower than the programmed value. "Overcurrent" lasting 
more than set value (1 to 9999) causes the channel to "trip". Output voltage will drop to zero either at the 
Ramp-down  rate  or  at  the  fastest  available  rate,  depending  on Power  Down  setting;  in both cases the 
channel is put in the off state. If trip= INFINITE, "overcurrent" lasts indefinitely. TRIP range: 0 ÷ 999.9s; 
1000 s = Infinite. Step = 0.1 s 

Voltage Ripple 

10Hz ÷ 1kHz 

Typ 

3 mVpp

 

5 mVpp

 

4 mVpp

 

Max 

5 mVpp

 

7 mVpp

 

8 mVpp

 

1kHz ÷ 20MHz 

Typ 

1 mVpp

 

3 mVpp

 

2 mVpp

 

Max 

3 mVpp

 

5 mVpp

 

5 mVpp

 

Accuracy 

Vmon vs. Vout  

± 0.02% ± 20mV

 

± 0.02% ± 100mV

 

± 0.02% ± 1V

 

Vset Vs. Vout  

± 0.02% ± 20mV

 

± 0.02% ± 100mV

 

± 0.02% ± 1V

 

Imon vs. Iout  

High range 

± 0.5% ± 50µA

 

± 0.2% ± 2µA

 

± 0.2% ± 1µA

 

± 0.2% ± 20nA

 

Low range 

± 0.5% ± 5µA

 

± 0.2% ± 200nA

 

± 0.2% ± 100nA

 

± 0.2% ± 2nA

 

Iset vs. Iout  

High range 

± 0.5% ± 50µA

 

± 0.2% ± 2µA

 

± 0.2% ± 1µA

 

± 0.2% ± 20nA

 

Low range 

± 0.5% ± 5µA

 

± 0.2% ± 200nA

 

± 0.2% ± 100nA

 

± 0.2% ± 2nA

 

Humidity range 

0 ÷ 80% non condensing 

Operating temperature 

0 ÷ 45°C 

Storage temperature 

-10 ÷ 70°C 

Altitude 

≤ 2000m a

bove sea level 

Vout / Temperature  coeff. 

±10 ppm/°C typ 

±50 ppm/°C typ 

Stability (24h; 1 hour warm up)

 

±0.01% /  full scale

 

±0.02% /  full scale

 

±0.01% /  full scale

 

±0.01% /  full scale

 

Long Term stability (1 week; 1h warmup)

 

±0.02% /  full scale

 

±0.04% /  full scale

 

±0.02% /  full scale

 

±0.02% /  full scale

 

Safety Standard – ROHS – Halogen free

 

RoHS compliant

 

EMC qualification

 

CEI EN 61326

 

 

 
 

1

 Validity range of specifications:  8031: 1% to 100%; 8032: 5% to 100%; 8033: 1% to 100%; 8034, 8034H: 5% to 100% 

2

 

Mixed

 boards have half of the channels with positive polarity and half with negative 

Summary of Contents for 803 Series

Page 1: ...803x Series 8 16 Ch Programmable HV Power Supplies Rev 10 19 December 2022...

Page 2: ...ible for damage caused by improper use of the device The manufacturer declines all responsibility for damage resulting from failure to comply with the instructions for use of the product The equipment...

Page 3: ...er ON 13 Local Control 14 BOARD Settings 14 Channel settings 15 Group Settings 16 Status Icon 18 Remote Control 19 Software tools 19 GECO2020 19 CAEN HV Wrapper 19 Power Supply Modules LabVIEW Instrum...

Page 4: ...ernating Current The following symbol may be reported in the present manual General warning statement The symbol could be followed by the following terms DANGER indicates a hazardous situation which i...

Page 5: ...ID ELECTRIC SHOCK OR FIRE HAZARD DO NOT OPERATE THIS PRODUCT WITH COVERS OR PANELS REMOVED CAUTION Do Not Operate in Wet Damp Conditions TO AVOID ELECTRIC SHOCK DO NOT OPERATE THIS PRODUCT IN WET OR D...

Page 6: ...ivered through BNC connectors other versions feature SHV s The HV output RAMP UP and RAMP DOWN rates may be selected independently for each channel Safety features include OVERVOLTAGE and UNDERVOLTAGE...

Page 7: ...Channels Voltage 100 240V Frequency 50 60Hz 0 8A rms max Fuse 2xT1A 6 3x32 250VAC Voltage 100 240V Frequency 50 60Hz 1 6A rms max Fuse 2xT2A 6 3x32 250VAC Voltage 100 240V Frequency 50 60Hz 1 6A rms m...

Page 8: ...CAEN Electronic Instrumentation 803x Series 8 16 Ch Programmable HV Power Supplies 8 Front panel...

Page 9: ...AEN Electronic Instrumentation 803x Series 8 16 Ch Programmable HV Power Supplies 9 External connections Local control section 2 8 LED Touch Screen Parameter and Mode setting Local settings monitoring...

Page 10: ...ferences can be used to provide a TTL compatible level on pin 2 and 3 In the first case externally supplied device the maximum allowed ratings are Maximum voltage between and 12V Maximum sink current...

Page 11: ...nnel Output 8032 33 34 34H series SHV RADIALL R317580 Impedance 50 Ohm Frequency range 0 2 GHz VSWR 1 20 0 3 F GHz plug and jack Test voltage 10kV DC 1mn unmated connectors Ratings 12kV DC 1mn mated p...

Page 12: ...he channel to trip Output voltage will drop to zero either at the Ramp down rate or at the fastest available rate depending on Power Down setting in both cases the channel is put in the off state If t...

Page 13: ...M mechanics The unit is an equipment for BUILDING IN install it in a NIM crate providing the required power supplies with forced cooling air flow then connect the HV Channel Output to the load R803x s...

Page 14: ...itor values are available also with remote control include Parameter Type Function Status Monitor Module power supply status HVMax Monitor Hardware Max Voltage settable via trimmer see p 10 HIMax Moni...

Page 15: ...left arrow to go back The right arrow allows to access to DHCP settings MAC Address 00 12 5e 00 07 07 DHCP Enabled If a DHCP Server is available then the module can be enabled or disabled as DHCP cli...

Page 16: ...ate V s Ramp Down Maximum High Voltage decrease rate V s Power Down Power Down mode after channel TRIP KILL or RAMP Trip Max time overcurrent allowed to last 1000 s Imon Range Current Monitor Zoom H o...

Page 17: ...VC OVP MAXV TRIP OVT OFF KILL ILK CAL_ERR single channels Vset High Voltage programmed value Volt Iset Current Limit programmed value A MaxV Absolute maximum High Voltage level that the channel can re...

Page 18: ...CAEN Electronic Instrumentation 803x Series 8 16 Ch Programmable HV Power Supplies 18 Status Icon Three types of Icon in the display status area indicate OK status WARNING status ALARM status...

Page 19: ...r software libraries For more info please visit www caen it products firmware software section Power Supply Modules LabVIEW Instrument Driver Power Supply Modules LabVIEW Instrument Driver is a set of...

Page 20: ...it 1 Flow control none Launch communication Type caen then enter Ethernet Launch the terminal emulator select port number 1470 Please note that line editing must be disabled prior to Ethernet access E...

Page 21: ...e to access the format EEPROM menu Channels settings By typing D on Terminal Board Menu it is possible to access channels settings To change one parameter point the parameter with the arrow keys and t...

Page 22: ...tation 803x Series 8 16 Ch Programmable HV Power Supplies 22 By typing G accesses Group Mode settings are extended to all channels I changes Interlock Mode see p 10 R resets alarm conditions see p 10...

Page 23: ...es 25 8 C Vout 50 4V If Text becomes 25 9 C Vout 50 4V If Text becomes 26 C Vout 50 5V The compensation is signed in fact if temperature decreases voltage is reduced If Text becomes 24 Vout 49 5V By i...

Page 24: ...se to a command of MON or INFO type it is not used in response to a command of SET type successful response to a command of SET type CMD OK CR LF successful response to a command of MON type CMD OK VA...

Page 25: ...efine for parameter type described by data format define NUMERIC 0 define ONOFF 1 define CHSTATUS 2 define BDSTATUS 3 define BINARY 4 define STRING 5 define ENUM 6 define READ 0 define WRITE 1 define...

Page 26: ...ON EN and one of these conditions is TRUE Bit 15 MAXV 1 VMON HVMAX set via trimmer Power On in local mode Switch from remote to local TRIP TRIP time value VTC Enable Temperature compensation Only for...

Page 27: ...s or flammable liquids on the screen Cleaning the air vents It is recommended to occasionally clean the air vents if present on all vented sides of the board Lint dust and other foreign matter can blo...

Page 28: ...ervice it is recommended to perform the following actions Detach all the signal input output cable Wrap the device in its protective packaging Insert the device in its packaging if present THE DEVICE...

Page 29: ...A Via Vetraia 11 55049 Viareggio Italy Phone 39 0584 388 398 Fax 39 0584 388 959 info caen it www caen it CAENspa INDIA Private Limited B205 BLDG42 B Wing Azad Nagar Sangam CHS Mhada Layout Azad Naga...

Reviews: