background image

CAEN

 

 

Electronic Instrumentation

 

 

803x Series 8-16 Ch. Programmable HV Power Supplies 

19 

Remote Control 

Module control can take place remotely, via USB or Ethernet; an external PC is required (supported OS: 
Windows 7 or later; Linux kernel Rel. 2.4 or later). The connection can be performed via VT emulators 
(such  as  Tera  Term),  or  via  one  of  the  available  tools  (see  below).  The  803x  Power  Supplies  family  is 
identified as 

SMART HV

 by the available software tools. Control option on the Touchscreen must be set 

to 

REMOTE

Software tools 

GECO2020 

CAEN GECO2020 is a graphical application that allows to control the 803x Power Supplies (and all other 
CAEN Power Supplies). Once the unit is correctly installed, download and install the 

GECO2020 software 

package related to your OS; follow the instructions in the GECO2020 User manual and the  unit will be 
ready to be operated. 

For more info, please visit www.caen.it (products>firmware/software section). 

CAEN HV Wrapper 

CAEN HV Wrapper is a library, available either as a set of ANSI C functions or LabVIEW

 

VI’s. Such set 

provides the software developer a unified software interface for the control of CAEN Power Supplies. This 
is a low-level application in which the writing of the Control SW is assigned to the user. It contains a generic 
software interface independent  by the Power Supply models and by the communication path used to 
exchange data with them. 

CAEN  HV  Wrapper  is  logically  located  between  a  higher-level  application,  such  as  GECO2020, and  the 
lower  layer  software  libraries. 

For  more  info,  please  visit  www.caen.it  (products>firmware/software 

section). 

Power Supply Modules LabVIEW Instrument Driver  

Power Supply Modules LabVIEW Instrument Driver is a set of VI’S, developed for LabVIEW 20

17 and later 

releases (LabVIEW

 is a Trademark of National Instruments Corp.), that allow to configure and monitor 

all parameters of remotely controlled CAEN Programmable HV Power Supply modules. Host PC shall run 
LabVIEW 2017 or later releases and NI-VISA Run-Time Engine 17. 

To install the Power Supply Modules LabVIEW Instrument Driver, g

o to CAEN web site in the “Software” 

area, download the Power Supply Modules LabVIEW Instrument Driver installation package and follow 
the Set Up instructions. 

PC connection 

USB installation 

Install the unit, as explained at p. 13 

Connect the Unit to the PC via the USB cable 

Turn ON the unit via the rear panel (R803x and 
DT803x) or NIM bin (N803x) switch  

Download and install the USB driver for your OS, 
available at the Unit page on the www.caen.it 
site

3

 

Now the Unit is ready for operation, upon 
installation of one of the available software tools 

Ethernet installation 

Install the unit, as explained at p. 13 

Connect the Ethernet port of the unit to the 
relevant port of the PC, using the 10BASE-T 
Ethernet cable 

Turn ON the unit via the rear panel (R803x and 
DT803x) or NIM bin (N803x) switch  

Now the unit is ready for operation, upon 
installation of one of the available software tools 

 

 
 

3

 Linux OS Users: the Unit is automatically recognised by Kernel Linux 2.6.9 and higher; unit name is assigned to serial 

port with name /dev/ttyACM[x], where [x] is device number; for example 1

st

 module connected is /dev/ttyACM0, 2

nd

 

module is /dev/ttyACM1 etc. 

Summary of Contents for 803 Series

Page 1: ...803x Series 8 16 Ch Programmable HV Power Supplies Rev 10 19 December 2022...

Page 2: ...ible for damage caused by improper use of the device The manufacturer declines all responsibility for damage resulting from failure to comply with the instructions for use of the product The equipment...

Page 3: ...er ON 13 Local Control 14 BOARD Settings 14 Channel settings 15 Group Settings 16 Status Icon 18 Remote Control 19 Software tools 19 GECO2020 19 CAEN HV Wrapper 19 Power Supply Modules LabVIEW Instrum...

Page 4: ...ernating Current The following symbol may be reported in the present manual General warning statement The symbol could be followed by the following terms DANGER indicates a hazardous situation which i...

Page 5: ...ID ELECTRIC SHOCK OR FIRE HAZARD DO NOT OPERATE THIS PRODUCT WITH COVERS OR PANELS REMOVED CAUTION Do Not Operate in Wet Damp Conditions TO AVOID ELECTRIC SHOCK DO NOT OPERATE THIS PRODUCT IN WET OR D...

Page 6: ...ivered through BNC connectors other versions feature SHV s The HV output RAMP UP and RAMP DOWN rates may be selected independently for each channel Safety features include OVERVOLTAGE and UNDERVOLTAGE...

Page 7: ...Channels Voltage 100 240V Frequency 50 60Hz 0 8A rms max Fuse 2xT1A 6 3x32 250VAC Voltage 100 240V Frequency 50 60Hz 1 6A rms max Fuse 2xT2A 6 3x32 250VAC Voltage 100 240V Frequency 50 60Hz 1 6A rms m...

Page 8: ...CAEN Electronic Instrumentation 803x Series 8 16 Ch Programmable HV Power Supplies 8 Front panel...

Page 9: ...AEN Electronic Instrumentation 803x Series 8 16 Ch Programmable HV Power Supplies 9 External connections Local control section 2 8 LED Touch Screen Parameter and Mode setting Local settings monitoring...

Page 10: ...ferences can be used to provide a TTL compatible level on pin 2 and 3 In the first case externally supplied device the maximum allowed ratings are Maximum voltage between and 12V Maximum sink current...

Page 11: ...nnel Output 8032 33 34 34H series SHV RADIALL R317580 Impedance 50 Ohm Frequency range 0 2 GHz VSWR 1 20 0 3 F GHz plug and jack Test voltage 10kV DC 1mn unmated connectors Ratings 12kV DC 1mn mated p...

Page 12: ...he channel to trip Output voltage will drop to zero either at the Ramp down rate or at the fastest available rate depending on Power Down setting in both cases the channel is put in the off state If t...

Page 13: ...M mechanics The unit is an equipment for BUILDING IN install it in a NIM crate providing the required power supplies with forced cooling air flow then connect the HV Channel Output to the load R803x s...

Page 14: ...itor values are available also with remote control include Parameter Type Function Status Monitor Module power supply status HVMax Monitor Hardware Max Voltage settable via trimmer see p 10 HIMax Moni...

Page 15: ...left arrow to go back The right arrow allows to access to DHCP settings MAC Address 00 12 5e 00 07 07 DHCP Enabled If a DHCP Server is available then the module can be enabled or disabled as DHCP cli...

Page 16: ...ate V s Ramp Down Maximum High Voltage decrease rate V s Power Down Power Down mode after channel TRIP KILL or RAMP Trip Max time overcurrent allowed to last 1000 s Imon Range Current Monitor Zoom H o...

Page 17: ...VC OVP MAXV TRIP OVT OFF KILL ILK CAL_ERR single channels Vset High Voltage programmed value Volt Iset Current Limit programmed value A MaxV Absolute maximum High Voltage level that the channel can re...

Page 18: ...CAEN Electronic Instrumentation 803x Series 8 16 Ch Programmable HV Power Supplies 18 Status Icon Three types of Icon in the display status area indicate OK status WARNING status ALARM status...

Page 19: ...r software libraries For more info please visit www caen it products firmware software section Power Supply Modules LabVIEW Instrument Driver Power Supply Modules LabVIEW Instrument Driver is a set of...

Page 20: ...it 1 Flow control none Launch communication Type caen then enter Ethernet Launch the terminal emulator select port number 1470 Please note that line editing must be disabled prior to Ethernet access E...

Page 21: ...e to access the format EEPROM menu Channels settings By typing D on Terminal Board Menu it is possible to access channels settings To change one parameter point the parameter with the arrow keys and t...

Page 22: ...tation 803x Series 8 16 Ch Programmable HV Power Supplies 22 By typing G accesses Group Mode settings are extended to all channels I changes Interlock Mode see p 10 R resets alarm conditions see p 10...

Page 23: ...es 25 8 C Vout 50 4V If Text becomes 25 9 C Vout 50 4V If Text becomes 26 C Vout 50 5V The compensation is signed in fact if temperature decreases voltage is reduced If Text becomes 24 Vout 49 5V By i...

Page 24: ...se to a command of MON or INFO type it is not used in response to a command of SET type successful response to a command of SET type CMD OK CR LF successful response to a command of MON type CMD OK VA...

Page 25: ...efine for parameter type described by data format define NUMERIC 0 define ONOFF 1 define CHSTATUS 2 define BDSTATUS 3 define BINARY 4 define STRING 5 define ENUM 6 define READ 0 define WRITE 1 define...

Page 26: ...ON EN and one of these conditions is TRUE Bit 15 MAXV 1 VMON HVMAX set via trimmer Power On in local mode Switch from remote to local TRIP TRIP time value VTC Enable Temperature compensation Only for...

Page 27: ...s or flammable liquids on the screen Cleaning the air vents It is recommended to occasionally clean the air vents if present on all vented sides of the board Lint dust and other foreign matter can blo...

Page 28: ...ervice it is recommended to perform the following actions Detach all the signal input output cable Wrap the device in its protective packaging Insert the device in its packaging if present THE DEVICE...

Page 29: ...A Via Vetraia 11 55049 Viareggio Italy Phone 39 0584 388 398 Fax 39 0584 388 959 info caen it www caen it CAENspa INDIA Private Limited B205 BLDG42 B Wing Azad Nagar Sangam CHS Mhada Layout Azad Naga...

Reviews: