REGUL8OR – User’s Manual
32
The inspection should confirm that there is no exterior damage to the digital
controller such as broken switch or connectors and all the panels and display are not
scratched or cracked. Keep all packing material until the inspection has been completed.
If damage is detected, compile the RMA form available from the CAEN ELS website.
2.3
Mounting
The REGUL8OR module can be used either as a desktop unit or as a rack-mount
device since the unit form factor is designed to be installed in a standard 19-inch cabinet
(1U space occupation).
This unit is natural air-cooled; ventilation holes are present
on the front and rear panel. Allow minimum 10 cm of
unrestricted air space at the front and the read of the unit.
Rack Mounting
The REGUL8OR units are designed to fit in a standard 19” equipment rack.
Use mounting rails to provide adequate support for the unit.
2.4
AC Input Power Connection
The AC line input connector on the rear panel is a standard IEC 60320 C20 male
inlet socket. Mains fuse holders are on the bottom part of the connector; required fuses
characteristics for all the models are (
T2AL250V
):
CAUTION
CAUTION
NO