Fit two AAA batteries in the battery compartment and close the cover.
Open the clamp as shown in the figure.
Insert a finger in the pulse oximeter.
Press the button on the front panel once.
The measurement is shown on the display.
1
2
3
4
5
Note:
Only Use alcohol to clean the rubber part of pulse oximeter in contact with the
patient’s finger (the rubber part in the device is medical rubber that does not
contain toxins and is not harmful to the skin).
Insert the finger in the
pulse oximeter with the
nail facing up.
MEASURING PROCEDURE
Equipment data update
period:
Data update period of
slower average is 8s
(Fig A).
The following details are
provided to disclose
actual performance
observed in the clinical
validation study
of healthy adult
volunteers. The ARMS
value analysis
statement and
Bland-Altman plot of
data is shown
as in the Fig B picture.
ARMS Value Analysis Statement
Item
90--100
80--<90
70--<80
#pts
78
66
63
Bias
1.02
0.40
-0.48
ARMS
1.66
1.46
1.93
Grafico a punti Bland-Altman
Fig A
Fig B
EN
EN
33
34
Summary of Contents for O2-easy
Page 1: ...ES IT DE FR PL PT RU Rev 5 del 19 12 2018 EN...
Page 83: ...164 163 1 2 3 O2 Easy RHb HbO2 660 905 SCHEMA DI FUNZIONAMENTO 1 2 CA MI 3 RU RU...
Page 84: ...30 8 SpO2 1 2 3 4 ESU 4 1 2 3 4 5 6 7 RU RU 165 166...
Page 85: ...3 8 9 10 11 12 13 14 15 16 17 SpO2 1 2 3 4 5 6 7 RU RU 167 168...
Page 88: ...1 2 3 A B RU RU 173 174...
Page 89: ...25 C 70 C 93 1 2 3 4 5 6 70 1 2 1 2 _ _ RU RU 175 176...
Page 90: ...15 10 70 70 2 EtO SpO2 PR SpO2 PR 1 2 1 2 1 2 1 2 RU RU 177 178...