TECHNICAL CHARACTERISTICS
Model
NEW ASKIR 20
Typology (MDD 93/42/EEC)
Medical Device Class II a
Classification UNI EN ISO 10079-1
HIGH VACUUM / LOW FLOW
Main Voltage
230 V ~ / 50 Hz
Power consuption
184 VA
Fuse
F 1 x 1.6A 250 V
Maximum suction aspiration (without jar)
-75kPa (- 0.75 bar)
Regolable from -75kPa (-0.75 Bar) to -10kPa (-0.10 Bar)
Maximum flow (without jar)
16 l/min
Weight
2.5 Kg
Dimension
350 x 210 x 180 mm
Functioning (to 35°C and 110 % operating voltage)
NON-STOP OPERATED
Working condition
Room temperature:
5 ÷ 35 °C
Room humidity percentage:
30 ÷ 75 % RH
Altitude:
0 ÷ 2000m s.l.m
Conservation condition
Room temperature:
- 40 ÷70 °C
Room humidity percentage:
10 ÷ 100% RH
SYMBOLS
Class 2 isolation equipment
CE marking in conformity with EC directive 93/42/CEE and subsequent changes
Manufacturer: CA-MI Srl - Via Ugo La Malfa n°31, 43010 Pilastro (PR) Italy
Warning, consult the instruction manual
Type B equipment
Fuse
To preserve in place coolness and dry land
Conservation temperature: - 40 ÷ 70° C
~
Alternate current
Hz
Mains frequency
I
ON
0
OFF
DEHP Phthalates (Suction catheter)
Please note tecnica specifications may vary upon the manufactures’s discretion!
Guidance and manufacturer’s declaration – electromagnetic Emissions
The surgical aspirator NEW ASKIR 20 is intended for use in the electromagnetic environment specified below.
The customer or the user of the surgical aspirator NEW ASKIR 20 should assure that it’s used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
Irradiated / Conducted emissions
CISPR11
Group 1
The surgical aspirator NEW ASKIR 20 only used RF energy only for its
internal functioning. Therefore, its RF emissions are very low and are
not cause interference in proximity of any Electronic appliances.
Irradiated / Conducted emissions CISPR11
Class [B]
The surgical aspirator NEW ASKIR 20 can be used in all environments,
including domestic and those connected directly to the public mains
distribution that supplies power to environments used for domestic
scopes.
Harmonic emissions IEC/EN 61000-3-2
Class [A]
Voltage fluctuations / flicker emissions
IEC/EN 61000-3-3
Complies
11