If the Signal Generator/Transmitter has successfully
completed the Self Test, ‘PASS’ will appear on the display.
The Signal Generator/Transmitter is ready for use as per normal.
It will not undertake another automatic daily Self Test that day.
If the Signal Generator/Transmitter has failed the Self Test,
‘FAIL’ will appear on the display and the Self Test warning
symbol will flash. The Self Test warning symbol will then remain
on the display.
After any Self Test fail, the Signal Generator/Transmitter
will automatically initiate a new Self Test when it is next
switched on.
If the Signal Generator/Transmitter fails a Self Test it should
be withdrawn from service. Contact C.Scope or a C.Scope
Authorised Service Centre for advice.