![Byond ResPlus C-20A CPAP User Manual Download Page 9](http://html1.mh-extra.com/html/byond/resplus-c-20a-cpap/resplus-c-20a-cpap_user-manual_2837202009.webp)
6
Heat Warning
Rx Only
Caution: Federal law restricts this device to sale by or on
the order of a licensed healthcare practitioner.
6. Device Operation
6.1 First Use
6.1.1 Place the device on a steady flat table where the settings are easy to reach and the
information on the display can be clearly seen by the user.
Maintain at least 2 inches between the device and the wall to ensure that the
air inlet is not obstructed.
When using the device, ensure that the surrounding air flows freely. Keep the
device away from any heating and cooling devices (air conditioning vents,
radiators, etc )
6.1.2 Place the air filter in the slot at the Air Inlet port.
6.1.3 Connect one end of the tubing into the Air Outlet.
6.1.4 Connect the mask into the other end of the tubing.
6.1.5 Connect the power adapter into the Power Port on the rear of the device.
6.1.6 Connect the device to mains power supply. The device will display the Standby
Interface
6.1.7 Connect the power supply. Press the On/Off Button to start the device.
The Connections should appear as shown in
Fig 6.1
below:
Fig.6.1
Put the device on a solid flat surface where the approach is not blocked, the
displays are easy to see, and the controls are convenient to access. Be sure to place the
device where it cannot fall easily. The device needs to be lower than the height of the
bed.
Do not place the device in or on a container where water may gather.
Be sure to take preventive measures in order to avoid water damage to the device.
When water is in the humidifier, do not move the device, so as to avoid water
entering into the device.
6.2 Daily Use
6.2.1 Assembling the Device