9
Fig.7.2.3a Fig.7.2.3b
(2) Add water through the Air Inlet port on top of the water tank. Make sure the water
level does not exceed the Max line.
Fig.7.2.3c
In winter months, be sure to add warm water, but no hotter than 35
℃
(
95
°F)
.
Do not fill the water tank above the Max line.
Turn off the humidifier when the water tank is empty. Turn the device off before
refilling water. Do not add water when the humidifier is in operation. Only add water
when the humidifier has cooled down.
Do not let any water enter the host unit.
8. Using the SpO2 Kit
The SpO2 Kit consists with a SpO2 Probe, Adapter, and Connector. The SpO2 Kit is
intended to be used for continuous, non-invasive functional arterial oxygen saturation
(SpO2) and pulse rate monitoring for adults weighing more than 40 kg (90 pounds).
The SpO2 Kit is immediately ready to use when it is connected to the main device via
the Communications Port. Attach the Oxygen Probe sensor to the user's index finger
(although other fingers can also be used).
The sampling frequency of the SpO2 signal is 50Hz, and the update frequency of the
device is 1Hz. The value of SpO2 is calculated by the average of the previous eight
pulse waveforms.
If the SpO2 Kit does not connect or is not performing correctly, the value of SpO2 will
not be displayed on the Main Interface.