GasAlertMicro
Calibration and Setting Alarm Setpoints
”.
If the span adjustment is an unusually large
(more than 20 %), the detector will advise:
note: Span
adjust unusually large - check that the
cal gas is correct.
Confirm that the calibration gas bottle you are using is
correct and that the span concentration value(s) in the
detector match the value of the calibration gas bottle.
See
the section
If the calibration adjustment is expected accept the
calibration.
Span Adjustment
Accept/Reject
If a span is rejected, the display advises:
Note: Span
rejected
and skips the span for the failed sensor(s).
Alarm Setpoints
Alarms may be set anywhere within the detection range for
the sensor. See the section “
” or set to zero for
off.
Factory alarm setpoints may vary by region. An example is
given below. Set as desired.
H
2
S
TWA: 10 ppm
STEL: 15 ppm
LOW: 10 ppm
HIGH: 15 ppm
CO
TWA: 35 ppm
STEL: 50 ppm
LOW: 35 ppm
HIGH: 200 ppm
O
2
LOW:
19.5%
HIGH:
23.5%
Combustible LOW: 10% LEL
HIGH: 20% LEL
SO
2
TWA: 10 ppm
STEL: 15 ppm
LOW: 10 ppm
HIGH: 15 ppm