GasAlertMicro
Users Manual
Table 13. Troubleshooting Tips (cont.)
Problem Possible
Cause Solution
Detector does not display normal
ambient gas reading after activation
self-test.
Sensor not stabilized.
Detector requires calibration.
Target gas is present.
→
Used sensor: wait 60 seconds
New sensor: wait 5 minutes
→
Calibrate detector. See the section
“
→
Detector is operating properly. Use
caution in suspect areas.
Detector does not respond to
pushbuttons.
Batteries are depleted.
Detector is performing operations that do
not require user input.
→
Replace batteries. See the section
“
→
Pushbutton operation restored
automatically when the operation
ends
Detector does not accurately measure
gas.
Detector requires calibration.
Detector is colder/hotter than ambient
gas.
Sensor filter is blocked.
→
Calibrate sensor. See the section
“
→
Allow the detector to acquire ambient
temperature before use.
→
Clean the sensor filter. See the
section “
40