ConneX1 Operator’s Manual
9
Activating/Deactivating the Detector
11.
If the last calibration failed, press to acknowledge. If the failure is not
acknowledged, the detector powers off.
12.
If the calibration is overdue, press to acknowledge. If calibration is not
overdue, proceed to step #
13
.
(Note: “ To start” alternates with “ To continue” every 4 seconds.)
If Force Calibration is enabled, a calibration must be completed before
entering normal operation. If a calibration is not performed, the detector
automatically powers off.
13.
If Cal Reminder is enabled, the screen displays the number of days before
a calibration should be performed.
If Cal Reminder is not enabled, the Calibration Interval displays. Proceed
to step #
14
.