Advanced settings
When a screen is no longer selected
or if its order has been changed, the general
order is automatically updated.
Fig. 3
3
:The "Graph" screen is no longer available
When a screen is selected again, it automatically moves to last place.
Fig. 3
4
:The "Graph" screen is available again, and in last place.
Setting the "Standard"
screen
7.7.4 Screen Settings
Press
[Reset Panel Param.]
to reset the control panel parameters.
7.7.5 Access - Password
From the "Settings" screen, press
[Config.] [Access/Password]
.
From the "Settings" screen, press [Config.] [Application Windows] [Std Window Parame-
ters].
Std-By Value
Display/Hide the leak rate display in Stand-by mode.
Inlet Pressure
Display/Hide the inlet pressure display.
Extra Pressure
Display/Hide the pressure display of for the cell or an external gauge.
The external gauge (at the customer's expense) is a gauge installed on the custom-
er's application, connected to the 37 pin I/O board.
Lower Display
Limit
Set the minimum value displayed for leak rate.
Leak rate not displayed if the value is less than the ’Lower Display Limit’ configured
value.
From the "Settings" screen, press [Config.] [Application Windows] [Std. Window Parame-
ters] [Leak Rate Bargraph].
Zoom on set
point
Activate zoom to set point.
Zoom to set point is used to display on the bargraph the reject set point centred on 2
decades.
Low Decade
Set the low decade for the bargraph display.
High Decade
Set the high decade for the bargraph display.
From the "Settings" screen, press [Config.] [Screen Settings].
Brightness
Set the brightness.
Contrast
Set the contrast.
Panel off
Activate the sleep mode screen.
The screen is in sleep mode when the back light goes off (black screen). The device
appears to be off, but this is not the case! Simply touching the screen reactivates the
display.
Paging Func.
Activate the Paging function.
When a
wireless
remote control (accessory) is used, the ’Paging’ function makes it
possible to easily find the remote if it is located within its field of use with the detec-
tor. When the function is activated, the remote emits a sound signal so it can be lo-
cated. To stop the sound signal, deactivate the Paging function.