Model 4420-V001
DIGISTANT
®
Page 86
5.2.1
SCPI Version
SCPI version: 1993.0
The SCPI version can be requested with the instruction SYSTem:VERSion? .
5.2.2
Measurement Instructions
CONFigure: CURRent
DESCRIPTION:
The device is switched to current measurement if it is in
the measurement mode or simultaneous transmission and
measurement mode.
SYNTAX:
CONFigure:CURRent
Parameters:
No parameters
REMARKS:
Enquiry form:
No enquiry form
Restrictions:
Instruction not permissible if - device in ramp mode,
- device in calibration mode,
- measurement in progress.
CONFigure:VOLTage
DESCRIPTION:
The device is switched to voltage measurement if it is in
the measurement mode or simultaneous transmission and
measurement mode.
SYNTAX: CONFigure:VOLTage
Parameters:
No parameters
REMARKS:
Enquiry form:
No enquiry form
Restrictions:
Instruction not permissible if - device in ramp mode,
- device in calibration mode,
- measurement in progress.
CONFigure:TEMPerature:FRTD
DESCRIPTION:
If the device is in the measurement mode or simultaneous
transmission and measurement mode, it is switched to RTD
measurement and the RTD type is selected
In the transmission mode, only the RTD type is selected.
SYNTAX:
CONFigure:TEMPerature:FRTD <parameter>
Parameters:
Pt100
-> set RTD to Pt100.
Ni100
-> set RTD to Ni100.
Pt200
-> set RTD to PT200.
Pt500
-> set RTD to Pt500.
Pt1000
-> set RTD to Pt1000.
REMARKS:
Enquiry form:
CONFigure:TEMPerature:FRTD?
5.2. SCPI Instruction list for the DIGISTANT
®
model 4420-V001