Model 4420-V001
DIGISTANT
®
Page 115
SYNTAX:
SOURce:VOLTage:RANGe <parameter>
Parameters:
Float value, with optional unit.
Value range:
0 to 11V
Valid units:
UV, MV, V, KV, MAV
Default unit:
V
REMARKS:
Enquiry form:
SOURce:VOLTage:RANGe?
On this enquiry, the voltage value is supplied in exponential form
(as volts) without an explicit unit.
Restrictions:
Instruction not permissible if - device in ramp mode
- device in calibration mode
- (only) measurement in progress
SOURce:RESistance:RANGe:AUTO
DESCRIPTION:
Automatic or manual range selection for RTD transmission is set
here.
SYNTAX:
SOURce:RESistance:RANGe:AUTO <parameter>
Parameters:
ON or 1 -> Automatic range selection
OFF or 0 -> Manual range selection
REMARKS:
Enquiry form:
SOURce:RESistance:RANGe:AUTO?
Parameters:
ON or 1 -> f automatic range selection is set
OFF or 0 -> If manual range selection is set
Restrictions:
Instruction not permissible if - device in ramp mode
- device in calibration mode
- (only) measurement in progress
SOURce:RESistance:RANGe
DESCRIPTION:
The device receives a temperature or resistance value from which
the RTD transmission range is determined in the case of manual
range selection. At the same time, manual range selection for RTD
transmission is set.
SYNTAX:
SOURce:RESistance:RANGe <parameter>
Parameters:
Float value, with optional unit.
Value range:
0 to 3000 K or 0 to 2kΩ,
Valid units:
C or CEL, F or FAR, K,
UOHM, MOHM, OHM, KOHM, MAOHM
Default unit:
Depending on the setting
REMARKS:
Enquiry form:
SOURce:RESistance:RANGe ?
On this enquiry, the temperature or resistance value is supplied as
a numerical value followed by the selected unit.
Restrictions:
Instruction not permissible if - device in ramp mode
- device in calibration mode
- (only) measurement in progress
SOURce:MODE
DESCRIPTION:
The transmission mode is set here: standard transmission, ramp 1
or ramp 2.
SYNTAX:
SOURce:MODE <parameter>