132
of 175
7.5.33 Read-out X-coordinates of current measurement curve
Slot
Index
Description
Value
Meaning of value
Type
Len R/W
132
0
Not possible
-
-
X
X
132
1…9
Reserved
-
-
X
X
132
10
Write access
:
If a curve is to be
read, it must be
prepared through a
write access before
the curve is first
read.
Read access
:
Index of the last
coordinate; if 0, there
is no curve
EVENT!
Writing any two
arbitrary bytes
initiates action
Integer value
0…4999
U16
U16
2
2
W_
R_
132
11…18
Reserved
-
-
X
X
132
19
Write access
:
Desired group of 200
coordinates. For
example, if
coordinates 600 ...
799 are to be
displayed, there must
be a 3 here. Query
the maximum
number of value
pairs under slot
132/10.
Read access
:
Group of 200
coordinates currently
displayed.
Integer value
0…24
Integer value
0…24
U16
U16
2
2
W_
R_
132
20
0. coordinate of
group of coordinates
Float value
Float according to
IEEE754
FLT
4
RO
132
21
1. coordinate of
group of coordinates
Float value
Float according to
IEEE754
FLT
4
RO
132
22
2. coordinate of
group of coordinates
Float value
Float according to
IEEE754
FLT
4
RO
132
23
3. coordinate of
group of coordinates
Float value
Float according to
IEEE754
FLT
4
RO
132
…
…
…
…
…
…
…
132
217
197. coordinate of
group of coordinates
Float value
Float according to
IEEE754
FLT
4
RO