WA-1150 and WA-1650 Wavemeter Operating Manual
3 Remote Communications
EXFO Burleigh Products Group Inc. (09042-M-00 Rev C)
3-15
SENSe Subsystem
The SENSe subsystem allows changes in how the instrument makes measurements and
provides information on the reference laser.
SENSe Subsystem Functions
:AVERage?
Description
Queries the number of readings being averaged for wavelength and power
values.
Query Response
{OFF | 2 | 4 | 8 | 16 | 32 | 64 | 128}
Examples
:SENS:AVER?
:AVERage {OFF | 2 | 4 | 8 | 16 | 32 | 64 | 128}
Description
Sets the number of readings being averaged for wavelength and power
values. It is the same as using the Average function through the display.
*RST value
OFF
Examples
:SENS:AVER OFF
:SENS:AVER 8
:MEDium?
Description
Queries the medium used for wavelength calculations.
Query Response
{AIR | VACUUM}
Examples
:SENS:MED?
:MEDium {AIR | VACuum}
Description
Sets the medium used for wavelength calculations. It is the same as using the
Medium function through the display.
*RST value
VAC
Examples
:SENS:MED AIR
:SENS:MED VAC
:POWeroffset?
Description
Queries the power offset being added to power values. It is the same as using
the Poffs function through the display. The power offsets are in units of
dBm.
Query Response
{OFF | 1 | 2 | 3 | ... | 19 | 20}
Examples
:SENS:POW?