background image

Photoplotter FilmStar-PLUS
Instructions to Use

Page 14 / 16

Here   you   do   not   need   any   connection   to   the 
plotter.
The software will create a plot-file (.fp2), which 
you can transfer to the plotter with a USB-stick.
Attention: the USB-Stick needs to be formatted 
in FAT16/FAT32 (NTFS is not supported!).
If you want to transfer the data via USB-stick, 
insert the stick containing the plot file 
into the USB-slot, switch on the plot-
ter and select "USB key". If you have 
multiple   plot   files   on   the   stick,   you 
can jump between the files back and 
forth with the arrow keys. Select a file 
by   pressing  

Enter

.   With  

X

  you   go 

back   to   the   welcome   screen.   The 
next steps in the display are self-ex-
plaining. Pressing  

X

,  

Return

  or  

Stop 

you will always return to the welcome 
screen. The stop button while plotting 
has  a delayed  reaction,  so the plot-
ting is not stopped by accidental con-
tact. If you want to cancel the plotting, 
press   the  

Stop

  button   for   a   few 

seconds.
After the start, it can take several seconds be-
fore  the  plotter  drum  reaches  its rated speed. 
Then the plot starts. The screen displays a pro-
gress   bar.   To   prevent   accidental   exposure   of 
the   film,   the   display   shuts   off   after   a   few 
seconds. To light the display again press the sil-
ver button.

Configuration:

When you install the software for the first time, 
the configuration will contain default values. To 
get proper layouts it is absolutely necessary to 
run the plotter with the machine specific para-
meters. To do so you can either go to 

Setup – 

Setup Table

 and type in the values from the printed data sheet, or, 

even better, go to 

Setup – Import Setup Table

 and import the file 

from the CD. This configuration file is named xxx-yy.phs (with xxx 
as serial number and yy as year, e.g. 707-13.phs).

Test light intensity:

 The light intensity settings are of course pro-

posal   values   only.   They  will  vary with   the   type   of   film   and   de-
veloper that you use, so you have to adapt them to your specific 
conditions by test. The settings are stored at menu 

Setup – Setup  

Table

 and there on register 

Laser intensity

. To find out the correct 

laser intensity the software comes with two integrated calibration 
functions. 

Bungard Elektronik GmbH & Co. KG, Rilkestraße 1, 51570 Windeck – Germany

Tel.: +49 2292 / 9 28 28-0, Fax: +49 2292 / 9 28 28-29

, E-mail: [email protected]

Summary of Contents for FilmStar-PLUS

Page 1: ...cs Windows routine for opening and saving of files higher resolutions in Gerber2Bitmap various cal ibration programs in Run_Filmstar 3 different plot speeds for either high precision or less time consumption Description The film is fixed on a rotating drum A highly focused red light laser diode is moving step wise alongside that drum driven by a precise stepper motor with worm and gear drive Of co...

Page 2: ...m 0 19685 mil 2 Finest space between 2 lines 10 µm 0 01 mm 0 3937 mil 3 Finest space between 2 areas 18 µm 0 018 mm 0 70866 mil 4 Finest inverse spot diameter transparent point in black area 25 µm 0 025 mm 0 98425 mil Safety Regulations Carefully read this instructions and especially the safety regulations before using the unit Also pay heed to the safety data sheets of the photographic chemicals ...

Page 3: ...processing temperature When handling chemicals wear personal protective equipment gowns gloves goggles The drum surface should be kept clean and dry To remove any remnants of the tape use a soft cloth moistened with a liquid detergent Do not use a spray The fixer contains silver and must be recycled Safety data sheets for the developer and fixer concen trates are available These are actually no da...

Page 4: ...d 209 While the Filmstar PLUS plots you can setup the photochemicals in the cuvette 1 mix developer 2 parts water one part developer 2 mix fixer 2 parts water one part fixer 3 the middle chamber fill with water After plotting has finished switch off the plotter remove the film in Dark Room condition and develop a minute and fix a minute Then rinse the film thoroughly and dry The result of your fir...

Page 5: ...main supply for the plotter and the safe light If you want to run the plotter directly from the PC then you need to establish a USB connection from the plotter inside of the Dark Room to the PC outside of the Dark Room We recommend vertical mounted tanks with cover for the chemical processing These need little in stallation footprint and the liquids have a small surface compared to their volume wh...

Page 6: ...ording to the film thickness There is another white arrow on the lower part of the drum housing that you should use for lateral positioning of the film After you fixed the upper edge of the sheet turn the drum by hand and sweep over the sheet so it goes tightly on the drum It is important that the sheet edges are parallel to the drum axis and that the upper and the lower sheet corners are facing e...

Page 7: ...plotting is completed or interrupted the drum stops to rotate and the laser head will return to its starting position Before you start a new plot please wait until the laser head has fully achieved its zero position on the left side of the drum Film Developing and Fixing In this chapter we assume that you are using the film and chemicals that we supplied and that you are treating the film in the D...

Page 8: ... any more Test Expose a piece of film for 2 seconds to normal daylight After max 1 minute of development at ambient temperature the film must be entirely black opaque If this is not the case the developer must be exchanged If you use the tank lid and take care not to pollute the developer it will have a life time of approx 6 months For frequent use the developer may be replenished with concentrate...

Page 9: ...m The plotting area shows the maximum film size that the plotter can handle Please set the value to what applies to your unit This is done in the menu Preference Plotting area In Preference Units you may set your preferred measuring unit metric inch A last but important setting that must be done only once is under P reference memory size Set this value according to the table below Computer RAM Set...

Page 10: ...h values the film will be structured more fine and the file size increases however enormous Since today s computers do well with large file sizes we recommend to choose a rather high resolution and reduce the resolution in the plotting process Annotation Standard Gerber If you have no Extended Gerber files you need to specify the measuring unit and the zero suppression for each Gerber file This is...

Page 11: ...pf is a simple converter that converts high resolution monochrome bitmaps into the FPF plotter format The bitmap format is univer sal but has the disadvantage that the files can be very large with a corresponding resolution For this reason we supply the converter Bit map2Fpf If your design program is not capable to pro duce Gerber files you can save your layout with the highest possible resolution...

Page 12: ...iles zoom into the image with F5 in steps 10 11 14 16 20 25 33 50 100 and zoom out by pressing F6 You can adjust the viewing section with the scroll bars on the right and below or click in an area in the small preview win dow View_FPF also contains a measur ing function e g to compare the bit map dimensions with the original Gerber data First select an appropriate magni fication then move the curs...

Page 13: ... the size of the software window To continue click on the triangular START button or press F5 In the next dialog please set the desired output resolution check the corresponding laser light intensity and chose whether you want to mirror or turn the image or make a negat ive plot The choice of plotting speed Slow Medium or Fast influ ences the drum speed and hence the plotting speed Depend ing on t...

Page 14: ...hen the plot starts The screen displays a pro gress bar To prevent accidental exposure of the film the display shuts off after a few seconds To light the display again press the sil ver button Configuration When you install the software for the first time the configuration will contain default values To get proper layouts it is absolutely necessary to run the plotter with the machine specific para...

Page 15: ... may measure the actual line spacing and enter the results into the columns Error on drum diameter and Drum lin earity As you will rarely need these micrometer range corrections we are not going more into details at this point If you need further assist ance please call our support We prefer the test grid which is generated with Tools Generate Test Grid to external produced test patterns because t...

Page 16: ...claims We cannot accept subsequent claims from damage or destruction of workpieces worked on in the ma chine because we have no knowledge or control over the operating conditions at your site This is valid in a general manner also for requirements from damage to articles buildings and persons as well as the environment We do not warrant that the function of the machine will meet the customer s req...

Reviews:

Related manuals for FilmStar-PLUS