background image

www.bullard.com

8

Maintaining the Fr

ee-Air Pump

Maintaining the Free-Air® Pump

Bullard EDP30 and EDP50 Free-Air Pumps consist of an electrically-
driven air pump with four carbon vanes. The vanes self-adjust as they 
wear and should last the following length of time if properly maintained:
 

EDP30 - 3,000 to 10,000 Hours 

 

EDP50 - 1,500 to 5,000 Hours

For the pump to operate at its optimum performance level, the following 
simple maintenance procedures must be performed routinely: 

Replace the Inlet and Outlet Filters Regularly

Dirty or clogged filters will restrict air flow to the respirators by causing a 
decrease in the pump’s outlet pressure. They may also decrease vane life 
and cause eventual pump overheating.
• Replace the inlet filter (Cat. No. 23611) at least once every 500 running 

hours or once per month, whichever is sooner.
The Inlet Filter keeps particulates larger than 25 microns out of the 
air supply and helps protect the pump’s carbon vanes.– Replace the 
CARBOFINE outlet filter cartridges (Cat. No. S17101) at least once every 
200 running hours or once per month, whichever is sooner.

The following are directions for replacing the outlet filter cartridge. These 
steps must be performed outside the 
contaminated work area:

a) Turn the pump off and disconnect 

it from the power source.

b) Disconnect the air supply hose 

attachment pod from the pump 
or the V60 extension hose by 
releasing the Kamlok clamps.

c) Remove the eight (8) wing nuts 

(with a wrench or pliers) and 
bolts holding the quick-discon-
nect coupler manifold to the pod 
(Figure 12).

d) Remove the filter cartridge plate 

from the pod.

e) Unscrew the wing nuts and re-

move the caps holding the outlet 
filter cartridges in their seats. 
Remove and discard the used 
filter cartridges (Figure 13).

f) Visually inspect the rubber 

gaskets on the filter cartridge 
plate for wear, cracks or tears 
(Figure 14).
If any sign of damage is 
detected, replace with a new 
gasket (Bullard Cat. Nos. 26701, 
filter side or 27161, cover side).
The old gasket will have to be scraped off the plate. Glue a new 

gasket onto the pod making sure the bolt holes are lined up with the 
gasket. 

g) Insert four (4) new filter cartridges in the seats and reassemble the 

caps and wing nuts to hold the cartridges in place.

h) Reassemble the pod. Tighten the wing nuts and bolts together firmly 

with a wrench or pliers, so that no air can escape between the three 
parts.

i) Mount the pod back onto the pump or V60 extension hose, locking the 

Kamlok clamps so that no air can escape.

j) Plug the pump back into its power source. 

Flush the Pump if Necessary

Should excessive dirt, sand, foreign particles, moisture or oil be permitted 
to enter the pump, the carbon vanes may become sluggish and the pump’s 
performance will deteriorate. This will result in decreased outlet pres-
sure or failure of the pump to operate.If the pump remains idle in a humid 
environment for a long period of time, rust film may build up in the pump’s 
chamber and/or rotor slots. This will result in decreased outlet pressure or 
a failure to operate at all. If the above occurs, the pump should be flushed 
with one of the following recommended safety solvents:
• Loctite Safety Solvent (Cat. No. S1791) 
• Inhibisol Safety Solvent 
• Dow Chemical Chlorothane

� 

WARNING

Never use kerosene or other combustible liquids or vapors with this pump. 
Their use may result in an explosion which may cause injury or death.

Directions for Flushing Pump:
1. Before flushing, disassemble and remove the following parts from the 

pump:
• Disconnect the respirator(s) air supply hose from the pod(s).
• Remove the pod or the V60 extension hose from the pump’s manifold.
• Remove the pump’s inlet filters.

2. With the pump running, pour several ounces of approved safety solvent 

into each of the pump’s air inlet ports. Repeat the flushing if necessary.

� 

WARNING

Run the pump for a sufficient time to purge all traces of the safety 
solvent before reconnecting the pods, inlet filters and air supply 

hoses and using the respirators. 

 

Avoid Running the Pump at Excessive Pressure

Avoid running the pump model EDP30 above 30 psig or the EDP50 above 
15 psig for any length of time. Running the pumps above the maximum 
recommended psig rating could cause motor damage and premature wear 
of the carbon vanes.
NEVER LUBRICATE THIS OIL-LESS AIR PUMP. THE MOTOR BEARINGS 
ARE GREASE PACKED AND SEALED. THEY REQUIRE NO FURTHER 
LUBRICATION.

Figure 12

Figure 14

Figure 13

Summary of Contents for EDP30

Page 1: ...require expensive carbon monoxide monitors high temperature alarms or airline filters No calibration is required Read all instructions before using your Free Air pump This manual provides detailed in...

Page 2: ...tact the respirator manufacturer or Bullard s Customer Service Department at 877 BULLARD or 1 859 234 6616 3 Supplied air respirators used with this pump MUST NOT be worn in any atmosphere immediately...

Page 3: ...EDP30 PUMP Pump s Output Air Pressure psig Pump s Output Air Volume cfm Maximum Recommended Mask Respirators Maximum Recommended Hood Respirators Motor Amp Load 5 32 Six Four 16 10 30 Six Four 18 15...

Page 4: ...pod s If V60 extension hose IS NOT required go to instruction 3 a Attach the V60 extension hose to the pump s manifold If extending only one pod connect the V60 hose to the manifold Kamlok coupler tha...

Page 5: ...ry Carbon Vane 4 vanes MAXIMUM AIR PRESSURE 30 psig 15 psig AIR FLOW 30 cfm at 10 psig 50 cfm at 10 psig PRESSURE RELIEF VALVES Adjustable Pre Set to open at 34 psig when adjusted fully closed Adjusta...

Page 6: ...f the respirator system being used to the quick disconnect coupler s or the pod s Figure 6 NOTE If your respirator s air supply hose approved by NIOSH as part of the respirator system being used will...

Page 7: ...respirators NOTE If the pod s are attached directly to the pump manifold the pressure gauges should always read within 5 of each other If the gauge on the manifold reads substantially higher than the...

Page 8: ...p with the gasket g Insert four 4 new filter cartridges in the seats and reassemble the caps and wing nuts to hold the cartridges in place h Reassemble the pod Tighten the wing nuts and bolts together...

Page 9: ...ough the respirator s air supply hose until it reaches ambient air temperature 2 If pressure relief valves are set too high reset to a lower pressure Be sure the outlet pressure remains in the approve...

Page 10: ...Gauge 0 15 psig S19683 Pressure Gauge 0 30 psig S19688 Pressure Gauge 0 60 psig S2084 Drive Belt Matched Pair EDP30 S2085 Drive Belt Matched Pair EDP50 S2089 Kamlok Adapter Male Pod S2090 Kamlok Coup...

Page 11: ...to factory inspection In warranty product of current design will be subjected to a rehandling charge less freight originally allowed All material must be shipped with transportation charges prepaid B...

Page 12: ...e outlet filter cartridge is multi layered with 40 micron size particles of activated carbon a sorbent bed with an exposed carbon surface of more than 15 000 square feet Vapor removal efficiency 90 95...

Reviews: