P2.3C, P2.4C

CAUTION

Hot surface

Burning hazard

According to the product type and operation conditions, the
temperature of the housing may exceed 50 °C during opera-
tion.

Depending on the conditions at the installation site it may
be necessary to provide these areas with appropriate warn-
ing signs.

5.1 Switching on the sample gas pump

Before switching on the device, ensure that:

• the hose and electrical connections are undamaged and

correctly installed,

• no parts of the sample gas pump have been dismantled

(e.g. cover),

• the gas inlet and outlet of the sample gas pump is not

shut,

• the preliminary pressure is under 0.5 bar,
• in the event of throttling under 150 l/h in continuous op-

eration, a bypass is available,

• the ambient parameters are complied with,
• information on rating plates is observed,
• the voltage and frequency of the motor correspond to

those of the network,

• the electrical connections are tightly fastened and the

monitoring devices have been connected and configured
correctly!

• air inlet openings and cooling surfaces are clean,
• protective measures have been carried out; earthing!
• the motor is secured correctly,
• the terminal box cover is closed and the cable entry

points have been properly sealed,

• the elastomer sprocket of the coupling (P2.4C only) is cor-

rectly installed and undamaged,

• depending on the operating mode, the necessary pro-

tective and monitoring devices are present and func-
tional (depending on the type of pump, e.g. motor circuit
breaker, manometer, flame arrestor, temperature monit-
oring).

When switching the sample gas pump on make sure that

• no abnormal sounds or vibrations occur.
• the flow rate is neither too low nor too high. This would

indicate a cracked bellow.

5.2 Operating the sample gas pump

The sample gas pump is intended exclusively for the pump-
ing of gaseous media. It is not suitable for liquids.

The sample gas pump should be operated without pre-com-
pression. A preliminary pressure of more than 0.5 bar is not
permitted. The gas outlet must not be shut.

The sample gas pump should be operated without pre-com-
pression. A preliminary pressure of more than 0.5 bar is not
permitted. The gas outlet must not be shut. The flow rate
must be at least 50 l/h. In the event of throttling under 150 l/

h in continuous operation, the flow rate must be regulated
via a bypass. In this case you should choose a version with by-
pass valve.

NOTICE

Extreme throttling reduces the life time of the bellow.

For pumps with an integrated bypass valve, the output
power can be adjusted. Do not expend a great amount of
power when turning the valve as otherwise the valve could
be damaged! The turning range of the valve is around seven
rotations.

NOTE: Read and observe the maintenance plan!

6 Maintenance

Always observe the applicable safety- and operating regula-
tions when performing any type of maintenance. Please refer
to the original operator’s manual on the included CD or on-
line at www.buehler-technologies.com for maintenance in-
formation.

7 Service and Repair

Please refer to the original operator’s manual on the included
CD or online at www.buehler-technologies.com for a detailed
description of the unit including information on
troubleshooting and repair.

13

Bühler Technologies GmbH

BX420022 ◦ 08/2018

Summary of Contents for P2.3C

Page 1: ...Ref 998BX420022 Analysentechnik Kurzanleitung Messgaspumpen deutsch 2 Brief Instructions Sample Gas Pumps english 8 Notice de montage Pompes de circulation fran ais 14 Gu a r pida Bombas de gases de m...

Page 2: ...tbe schreibung und Betrieb und Bedienung sowie die Angaben der Datenbl tter hinsichtlich spezifischen Verwendungs zwecks vorhandener Werkstoffkombinationen sowie Druck und Temperaturgrenzen Bei einer...

Page 3: ...treten hoher Gastemperaturen durch adiabatische Kompression ist m glich und vom Anwender zu pr fen Achten Sie auf die Einhaltung der zul ssigen Daten und Ein satzbedingungen insbesondere auf die zul s...

Page 4: ...ungen etc sein Verwenden Sie niemals Ger te mit offensichtlichen Besch digungen VORSICHT Verwenden Sie geeignetes Werkzeug In bereinstimmung mit der DIN EN 1127 1 unterliegt die Handhabung und Auswahl...

Page 5: ...en Anschl ssen versehen Vergleichen Sie die Artikelnummer auf dem Typen schild mit der Artikelnummerstruktur im Kapitel Einleitung Vermeiden Sie Mischinstallationen d h Rohrleitungen an Kunststoffk rp...

Page 6: ...dass der Pumpenmotor die korrekte Spannung und Frequenz hat Spannungstoleranz 5 und Frequenzto leranz 2 Die Zuleitungs sowie Erdungsquerschnitte sind der Bemes sungsstromst rke anzupassen Verwenden S...

Page 7: ...der Elastomer Zahnkranz der Kupplung nur P2 4C rich tig montiert und nicht besch digt ist die abh ngig vom Betrieb notwendigen Schutz und berwachungsvorrichtungen vorhanden und funktions f hig sind j...

Page 8: ...d use existing material combin ations as well as pressure and temperature limits in the data sheets When installed outdoors ensure adequate protection from the weather see chapter Requirements for the...

Page 9: ...ust automatically shut down the sample gas pump should the temperature exceed the limits DANGER DANGER Explosion danger in case of high temperat ures Temperature of the device depends on the medium te...

Page 10: ...uld result in equipment damage CAUTION Avoid vibrations and resonances The operator is responsible to mount the pump in a way that vibrations and resonance do not cause premature failure res ulting in...

Page 11: ...es are tight 4 4 1 Monitoring the sample gas pump NOTICE When following preventive maintenance according to the maintenance plan a crack in the bellows is a rare malfunc tion but cannot be completely...

Page 12: ...he operation site must cor respond to all the ratings plate information 5 Operation and control NOTICE The device must not be operated beyond its specifications DANGER Danger of explosions danger of p...

Page 13: ...When switching the sample gas pump on make sure that no abnormal sounds or vibrations occur the flow rate is neither too low nor too high This would indicate a cracked bellow 5 2 Operating the sample...

Page 14: ...n du produit et fonctionnement et utilisation ainsi que les informations des fiches techniques concernant les usages sp cifiques les com binaisons de mat riaux pr sentes ainsi que les limites de press...

Page 15: ...anch it des tuyaux DANGER Compression adiabatique risque d explosion L apparition de temp ratures de gaz lev es pour cause de compression adiabatique est possible et doit tre v rifi e par l utilisateu...

Page 16: ...ez les trois vis cruciformes du couvercle de bo tier et enlevez celui ci voir im 1 im 2 Le m canisme d en tra nement est maintenant visible ATTENTION Risque d crasement Risque d crasement des doigts N...

Page 17: ...ez que la pompe de circulation est compl te Il faut encore 6 vis M6 avec crous de longueur ad quate pour le montage Sur tous les types de pompes la t te de pompe peut unique ment tre orient e 0 et 180...

Page 18: ...une surveillance CMT CMT Concentration maximale sur le lieu de travail sur le lieu de mise en place Manom tre de sous pression Pompe de gaz de mesure Manom tre D bitm tre Unit de coupure Fig 1 Exemple...

Page 19: ...x gazeux peut entra ner dans le soufflet le corps de la pompe une charge lectrostatique entra nant un risque d inflammation surface projet e dans le soufflet le corps de pompe 15 cm2 DANGER Risque d e...

Page 20: ...e r duction diminue la dur e de vie du soufflet Sur les pompes avec vanne de d rivation int gr e les perfor mances de refoulement peuvent tre r gul es N exercez pas une force excessive pour tourner la...

Page 21: ...los cap tulos Descripci n del producto y Uso y funcionamiento as como las indicaciones descritas en las hojas de datos en re laci n al uso previsto espec fico las combinaciones de mate riales disponi...

Page 22: ...diab tica peligro de explosi n El usuario debe comprobar si se producen altas temperaturas a causa de una compresi n adiab tica Aseg rese de que se cumplen los datos aportados y las condi ciones de in...

Page 23: ...No utilice nunca dispositivos en los que se aprecien desperfectos CUIDADO Utilice herramientas espec ficas De acuerdo con la norma DIN EN 1127 1 la utilizaci n y selec ci n de herramientas espec fica...

Page 24: ...para evitar que la condensaci n se escape y no se acumule en la bomba o el conducto 4 4 Conexi n de las tuber as de gas Las bombas han de estar previstas de las conexiones que ha elegido Compare el n...

Page 25: ...gurarse contra un sobrecalentamiento no autorizado a trav s de un protector contra sobrecarga rel t rmico Tener en cuenta la corriente asignada para la configuraci n del interruptor de protecci n v as...

Page 26: ...as La tensi n y la frecuencia del motor coincidan con los va lores de red Las conexiones el ctricas est n fijadas firmemente y los dispositivos de monitorizaci n est n conectados y confi gurados debid...

Page 27: ...del dispositivo y recomendaciones recomendaciones en cuanto al manteni miento y la soluci n de problemas consulte las instrucciones originales que se pueden encontrar en el CD que se incluye y en Inte...

Page 28: ...tschland 49 0 2102 4989 0 49 0 21 02 49 89 20 1 1 P2 3C P2 4C P2 3C P2 4C II 3G c IIB T4 X P2 3C P2 4C IIA IIB 80 1 2 P2 3C P2 4C 1 x 1 x 4 x 1 x 1 x 1 4301 1 x 1 x 1 x 2 a b c d a b c T4 T4 80 3 20 C...

Page 29: ...P2 3C P2 4C 1 2 1 2 4 DIN EN 1127 1 4 1 20 C 50 C 1000 m NN d 4 1 1 4 2 P2 3C P2 3C P2 4C P2 4C 42 011 14 3 6 x M6 0 180 4 3 4 4 PTFE 29 B hler Technologies GmbH BX420022 08 2018...

Page 30: ...P2 3C P2 4C In Out 4 4 1 UEL 4 4 1 1 4 4 1 2 1 2 1 2 UEL LEL MAK MAK 1 4 5 5 2 1 5 2 DIN VDE 0100 5 a b c d T4 15 cm2 P2 x Atex 1 30 B hler Technologies GmbH BX420022 08 2018...

Page 31: ...P2 3C P2 4C 80 50 C 5 1 0 5 150 l h P2 4C 5 2 0 5 0 5 50 l h 150 l h 7 6 CD www buehler technologies com 7 CD www buehler technologies com 31 B hler Technologies GmbH BX420022 08 2018...

Page 32: ...GmbH Harkortstra e 29 40880 Ratingen Deutschland 49 0 21 02 49 89 0 49 0 21 02 49 89 20 1 1 P2 3C P2 4C P2 3C P2 4C II 3G c IIB T4 X P2 3C P2 4C IIA IIB 80 1 2 P2 3C P2 4C 1 x 1 4 x 1 x 1 1 4301 1 1...

Page 33: ...P2 3C P2 4C a b c d a b c d a b c T4 T4 80 33 B hler Technologies GmbH BX420022 08 2018...

Page 34: ...P2 3C P2 4C 50 C 3 20 C 40 C 4 F 104 F 1 5 M 20 C 68 F 1 2 1 2 4 DIN EN 1127 1 4 1 20 C 50 C 1000 34 B hler Technologies GmbH BX420022 08 2018...

Page 35: ...P2 3C P2 4C d 4 1 1 4 2 P2 3C P2 3C P2 4C P2 4C 42 011 14 3 6 x M6 0 180 4 3 4 4 PTFE In Inlet Out Outlet 4 4 1 OEG 35 B hler Technologies GmbH BX420022 08 2018...

Page 36: ...P2 3C P2 4C 4 4 1 1 4 4 1 2 1 2 1 2 OEG UEG MAK 1 4 5 5 2 1 5 2 DIN VDE 0100 36 B hler Technologies GmbH BX420022 08 2018...

Page 37: ...P2 3C P2 4C 5 a b c d T4 2 15 2 80 50 C 5 1 0 5 150 37 B hler Technologies GmbH BX420022 08 2018...

Page 38: ...P2 3C P2 4C P2 4C 5 2 0 5 0 5 50 150 7 6 www buehler technologies com 7 www buehler technologies com 38 B hler Technologies GmbH BX420022 08 2018...

Reviews: