TC-MIDI (+)
5.8 Modbus Register
Description
FC
Address
Access Data type
Default
Min
Max
Selection
Resolution Unit
Block temperature measurement
3
2000
R
Float
-
-
-
-
0.5
°C
Block temperature status
3
2002
R
Uint32
-
-
-
Bit 0 := Error
Bit 1..15 := Reserved
Bit 16:= Sensor not calibrated
Bit 17:= Initialization/measurement invalid
Bit 18 := Stabilisation phase
Bit 19:= Load limit reached
Bit 20:= Measurement outside target range
Bit 21..31 := Not used
-
-
Ambient temperature measure-
ment
3
2004
R
Float
-
-
-
-
0.5
°C
Ambient temperature status
3
2006
R
Uint32
-
-
-
Bit 0 := Error
Bit 1..15 := Reserved
Bit 16 := Sensor not calibrated
Bit 17 := Initialization / measurement invalid
Bit 18..31 := Not used
-
-
Block temperature setpoint
3, 16
5000
R/W
Float
5.0
3.0
20.0
-
0.5
°C
Positive alarm tolerance setpoint
3, 16
5002
R/W
Float
3.0
1.0
7.0
-
1.0
K
Negative alarm tolerance setpoint 3, 16
5004
R/W
Float
-3.0
-3.0
-1.0
-
1.0
K
Temperature difference Delta-T
3, 16
5008
R/W
Float
-15.0
-30.0
0.0
1.0
K
Enable/disable Delta T
3, 16
9001
R/W
Uint16
0
-
-
0 := Normal mode
1:= Delta-T control
1
-
Moisture detector
signal store error
3, 16
9002
R/W
Uint16
2
-
-
1 := No
2 := Yes
-
-
Moisture alarm
signal store
3, 16
9003
R/W
Uint16
1
-
-
1 := No
2 := Yes
-
-
Sensitivity moisture detector 1
3, 16
9004
R/W
Uint16
1 (with FF)
2 (without FF)
-
-
0 := Sensitivity low
1 := Sensitivity high
2 := Moisture detector inactive
-
-
Sensitivity moisture detector 2
3, 16
9005
R/W
Uint16
1 (with FF)
2 (without FF)
-
-
0 := Sensitivity low
1 := Sensitivity high
2 := Moisture detector inactive
-
-
Display unit selection
3, 16
9006
R/W
Uint16
1
-
-
1 := °C
2 := °F
-
-
Heat exchanger type selection
3, 16
9007
R/W
Uint16
2
-
-
2 := Steel
3 := Glass
4 := PVDF
-
-
27
Bühler Technologies GmbH
BE440016 ◦ 07/2022