GAS 222.20 Ex1

Passing through gases

Flammable gases above the UEL (upper explosion limit) may
only be blown back with inert gases. Flammable gases from
25 % LEL (lower explosion limit) and up to the LEL may be
blown back provided the operator ensures the blown back gas
is not and cannot be explosive. For safety reasons we recom-
mend only using inert gases in these cases as well.

Blowing back explosive atmospheres (range from LEL to UEL)
with the probes is prohibited due to possible adiabatic com-
pression (high blowback pressure against contaminated fil-
ter). The operator is responsible for compliance with these
conditions taking into account his risk assessment.

Limitations categories/zones due to accessories

Please note, the accessories used may limit the approved ap-
plications of the probes.

Please note the following table:

ATEX +

IECEX

Only ATEX

GAS 222
types

with accessories:

Gas

Dust

Gas and

dust

(separate

zones)

Sampling zone/operating

zone

11 Ex1, 
21 Ex1, 
30 Ex1, 
31 Ex1, 
35 Ex1,
35-U Ex1

Pressure vessel PAV 01
(item no. 46222PAV
and accessories)

Zone1***/

Zone 1

Zone 20/

Zone 21

Zone 20/

Zone 1

11 Ex1, 
21 Ex1, 
30 Ex1, 
31 Ex1

Ceramic inlet filter*
(item no.:46
46222307F)

Zone 2/

Zone 1

Zone 20/

Zone 21

Zone 20/

Zone 1

or

Zone 2 /

Zone 21

11 Ex1, 
20 Ex1, 
21 Ex1

Ceramic outlet filter*
(item no. 46
46222026P)

Zone 2/

Zone 1

Zone 20/

Zone 21

Zone 20/

Zone 1

or

Zone 2 /

Zone 21

11 Ex1, 
20 Ex1, 
21 Ex1

Sampling tubes
(item no.: 46222001,
462220011, 46222006,
46222004, 46222016,
46222017, 46222018

Zone 0/

Zone 1

No zone/

Zone 21

Zone 0/

Zone 21

11 Ex1, 
20 Ex1, 
21 Ex1

Ceramic sampling
tubes**
(item no.:
4622200205,
4622200210,
4622200215)

Zone 2/

Zone 1

No zone/

Zone 21

Zone 2/

Zone 21

* Accessories not suitable for sampling highly ignitable dusts
with a minimum ignition energy (MIE) of < 3 mJ.

** When sampling gas from Zone 2, ceramic sampling tubes
may only be used if application- and process-related intensive
electrostatic charges are ruled out.

*** Blowback of explosive atmosphere/gases is prohibited.

1.2 Dust Atmosphere Requirements

The sample gas probe may only be used in areas with explos-
ive dust atmosphere if the smouldering temperatures of the
dust layers and the ignition temperatures of the dust atmo-
sphere is higher than the following temperatures.

Probe temperature class

T80 °C T120 °C T130 °C T175 °C T226 °C T300 °C

Smoul-
dering
temper-
ature
for
max. 5
mm
dust
layer

≥ 155 °C ≥ 195 °C ≥ 205 °C ≥ 250 °C ≥ 301 °C ≥ 375 °C

Dust at-
mo-
sphere
ignition
temper-
ature

≥ 120 °C ≥ 180 °C ≥ 195 °C ≥ 263 °C ≥ 339 °C ≥ 450 °C

The basis for safety factors is EN 1127-1.

1.3 Type Plate

Example:

0158

GAS 222.20 Ex1    
000092697 462222010149420013111 001

Manufacturer and address

Model designation

Order no., item no.

Electrical supply

Type examination certificate no.

230V; valve/клапан 24V UC

Bühler Technologies GmbH
Harkortstr. 29     D-40880 Ratingen

IBExU17ATEX1088X

IECEx IBE 17.0024X
Ex eb mb IIC T4 Gb

Explosion protection marking

II 2G

IECEx certificate number:

Ex eb mb IIC T4 Gb

ATEX certificate number:

Explosion protection marking

1.4 Scope of delivery

• 1 x Sample gas probe
• 1 x Flange gasket and screws
• Product documentation
• Connection and mounting accessories (only optional)

2 Safety instructions

2.1 Important Information

This unit may only be used if:

• the product is being used under the conditions described

in the operating- and installation instructions, used ac-
cording to the nameplate and for applications for which it
is intended. any unauthorized modifications to the device
will void the warranty provided by Bühler Technologies
GmbH,

• the limits in the data sheet and the instructions must be

observed,

11

Bühler Technologies GmbH

BX460041 ◦ 03/2021

Summary of Contents for GAS 222.20 Ex1

Page 1: ...60041 Version 03 2021 Ref 998BX460041 Analysentechnik Kurzanleitung Sonden deutsch 2 Brief Instructions Probes english 10 Notice de montage Sondes fran ais 16 Gu a r pida Sondas espa ol 23 chinese sim...

Page 2: ...ner m glichen Explo sionsgefahr nicht erlaubt Prozessgase oder Prozessgasgemi sche d rfen keine Feststoffe enthalten die in Kombination mit den Materialien der Sonden inklusive Filter und Entnah meroh...

Page 3: ...rtungsbereich des Betreibers unter Zuhilfe nahme seiner Risikobetrachtung Einschr nkungen der Kategorien Zonen durch Zubeh r Es ist zu beachten dass je nach verwendetem Zubeh r der zu gelassene Einsat...

Page 4: ...Haftung durch die B hler Techno logies GmbH ausgeschlossen die im Datenblatt und der Anleitung angegebenen Grenz werte eingehalten werden das Griffst ck samt O Ring mit geeignetem Umgebung stemperatur...

Page 5: ...Beachtung der Vorschrift IEC EN 60079 14 Zus tzliche nationale Bestimmungen bez glich Inbetrieb nahme Betrieb Wartung Instandhaltung und Entsorgung sind einzuhalten GEFAHR Elektrische Spannung Gefahr...

Page 6: ...chichten m ssen oberhalb der ge kennzeichneten Oberfl chentemperatur liegen und um einen Sicherheitsfaktor erg nzt sein siehe auch Abschnitt Anfor derungen an die Staubatmosph re 2 3 Umgebungstemperat...

Page 7: ...al muss den Anwendungsvoraussetzungen entsprechen und wetterfest sein 4 6 Anschluss der Gasleitung Die Messgasleitung ist mittels geeigneter Verschraubung sorgf ltig und fachgerecht anzuschlie en Dies...

Page 8: ...utzvor schriften Der Hersteller des Heizbandes empfiehlt in re gelm igen Abst nden den Isolationswiderstand des Heizbandes zu messen Der Isolationswiderstand ist dabei zwischen kurzgeschlossenen Kupfe...

Page 9: ...usf hrliche Beschreibung des Ger tes mit Hinweisen zur Fehlersuche und Reparatur finden Sie in der Originalbe triebsanleitung auf der beigef gten CD oder im Internet un ter www buehler technologies co...

Page 10: ...em perature class for zone 0 1 for zone 1 ATEX II 1G 2G Ex db1 eb mb2 IIC T5 T6 T1 T2 Ga Gb ATEX II 2G Ex db1 eb mb2 IIC T6 T2 Gb IECEx Ex db1 eb mb2 IIC T5 T6 T1 T2 Ga Gb IECEx Ex db1 eb mb2 IIC T6 T...

Page 11: ...ng highly ignitable dusts with a minimum ignition energy MIE of 3 mJ When sampling gas from Zone 2 ceramic sampling tubes may only be used if application and process related intensive electrostatic ch...

Page 12: ...andatory maintenance is performed Legal regulations are observed during disposal compliance with national installation regulations NOTICE Accessories may limit critical operating parameters of the bas...

Page 13: ...may be limited based on the version Please note the Ambient temperature ranges under Technical Data in the full operating instructions 2 4 Permissible gas inlet temperature The permissible gas inlet t...

Page 14: ...ssible screw connection Shorten the connection pipe for the sample gas line as much as possible To do so remove the insulation around the sample gas line This is done by loosening the fixing bolts CAU...

Page 15: ...lowback 5 Operation and Control NOTICE The device must not be operated beyond its specifications NOTICE The weather hood must be closed during operation DANGER Explosion hazard due to electrostatic di...

Page 16: ...sonde de pr l vement de gaz n est pas autoris en raison d un risque d explosion Les gaz ou m langes de gaz in dustriels ne doivent pas contenir de mati res solides qui com bin es aux mat riaux des son...

Page 17: ...t de l exploitant sur la base de son va luation des risques Limitations des cat gories zones du fait des accessoires Sachez que en fonction des accessoires utilis s le domaine d application autoris de...

Page 18: ...ambiante appropri e et filtre le cas ch ant les dispositifs de surveillance le dispositif de protection sont correctement raccord s les travaux de maintenance et de r paration non d crits dans ce mod...

Page 19: ...montage installation et main tenance ne doivent tre r alis s qu en absence d atmosph re explosive DANGER Utilisation dans des zones risque d explosion Les atmosph res charg es de gaz ou de poussi res...

Page 20: ...une appartenance un groupe d explosion des classes de temp rature ou des r sistances chimiques diff rentes de ceux de l appareil de base Int grez toujours toutes les donn es techniques des instruc tio...

Page 21: ...s s choisis soient plut t courts Raccourcissez le tube de raccordement de la conduite de gaz de mesure autant que possible Retirez pour cela le re v tement isolant voire les mors isolants dans la zone...

Page 22: ...bo te de connexion voir sch ma en annexe pour l af fectation des contacts des bornes pour le branchement des lectrovannes en cas de r trolavage automatique son pr vues 5 Fonctionnement et utilisation...

Page 23: ...gas de muestreo debido al posible riesgo de explosi n Los gases de proceso o las mezclas de gases de proceso no deben contener ning n s lido que en combinaci n con los materiales de las sondas incluid...

Page 24: ...del operador con ayuda de su evaluaci n de riesgos Limitaciones de categor as zonas mediante accesorios Debe tenerse en cuenta que dependiendo de los accesorios utilizados el rango de aplicaci n permi...

Page 25: ...La construcci n de instalaciones el ctricas en zonas con riesgo de explosi n requiere el cumplimiento de la norma IEC EN 60079 14 Tambi n deber n respetarse las restantes normativas na cionales relati...

Page 26: ...ones No se permite la extracci n de gases mezclas de gases o polvos que puedan ser explosivos en presencia de aire Acumulaci n de electricidad est tica formaci n de chispas Los equipos el ctricos solo...

Page 27: ...e especial atenci n a la longitud de ampliaci n del conducto de la sonda En caso de que la sonda se lleve al lugar de montaje por pie zas deber volver a montarse en primer lugar 4 2 Montaje PELIGRO Pe...

Page 28: ...incorrecta puede destrozar el dispositivo Comprobar en la conexi n que la tensi n de red sea la correc ta de acuerdo con la placa indicadora CUIDADO Da os en el dispositivo Deterioro del cable No da e...

Page 29: ...control necesarios est n disponibles y operativos por ej barrera contra llamas las salidas y entradas de gas de la sonda de gas de mues treo no est n bloqueadas se cumplen los par metros de entorno la...

Page 30: ...x db1 eb mb2 IIC T5 T6 T1 T2 Ga Gb IECEx Ex db1 eb mb2 IIC T6 T2 Gb 0 21 20 1 ATEX II 1G 2D Ex db1 eb mb2 llC T5 T1 Ga Ex tb mb2 lllC T80 C T226 C Db ATEX II 1D 2G Ex ta lllC T120 C T300 C Da Ex db1 e...

Page 31: ...0 1 21 0 21 11 Ex1 20 Ex1 21 Ex1 4622200205 4622200210 4622200215 2 1 21 2 21 MIE 3 mJ 2 1 2 T80 C T120 C T130 C T175 C T226 C T300 C 5 mm 155 C 195 C 205 C 250 C 301 C 375 C 120 C 180 C 195 C 263 C 3...

Page 32: ...GAS 222 20 Ex1 IEC EN 60079 14 a b c d a b c a b c M3 20 21 22 5 mm 2 3 2 4 3 20 C 50 C 4 F bis 122 F 4 32 B hler Technologies GmbH BX460041 03 2021...

Page 33: ...222 20 Ex1 4 1 4 2 4 3 4 4 O O O 90 4 5 4 6 GAS 222 1 DN65 PN6 DN3 150 G3 4 NPT 1 4 1 6 mm 1 4 1 1 NPT 1 4 4 6 1 6 mm 1 4 6mm 1 4 4 7 100 C 212 F 100 C 212 F 33 B hler Technologies GmbH BX460041 03 2...

Page 34: ...GAS 222 20 Ex1 4 7 1 4 7 2 30 mA RCD 2500 V DC 10 MOhm 5 5 1 O 6 CD www buehler technologies com 7 CD www buehler technologies com 34 B hler Technologies GmbH BX460041 03 2021...

Page 35: ...T6 T2 Gb IECEx Ex db1 eb mb2 IIC T5 T6 T1 T2 Ga Gb IECEx Ex db1 eb mb2 IIC T6 T2 Gb 0 21 20 1 ATEX II 1G 2D Ex db1 eb mb2 llC T5 T1 Ga Ex tb mb2 lllC T80 C T226 C Db ATEX II 1D 2G Ex ta lllC T120 C T...

Page 36: ...Ex1 30 Ex1 31 Ex1 46222307 46222307F 2 1 20 21 20 1 2 21 11 Ex1 20 Ex1 21 Ex1 46222026 46222026P 2 1 20 21 20 1 2 21 11 Ex1 20 Ex1 21 Ex1 46222001 462220011 46222006 46222004 46222016 46222017 4622201...

Page 37: ...58 GAS 222 20 Ex1 000092697 462222010149420013111 001 230V valve 24V UC B hler Technologies GmbH Harkortstr 29 D 40880 Ratingen IBExU17ATEX1088X IECEx IBE 17 0024X Ex eb mb IIC T4 Gb II 2G IECEx Ex eb...

Page 38: ...GAS 222 20 Ex1 IEC EN 60079 14 a b c d a b c a b c M3 38 B hler Technologies GmbH BX460041 03 2021...

Page 39: ...GAS 222 20 Ex1 20 21 22 5 2 3 2 4 3 20 C 50 C 4 F 122 F 4 4 1 4 2 4 3 39 B hler Technologies GmbH BX460041 03 2021...

Page 40: ...GAS 222 20 Ex1 4 4 90 4 5 4 6 GAS 222 1 DN65 PN6 DN3 150 G3 4 NPT 1 4 1 6 1 4 1 1 NPT 1 4 4 6 1 6 1 4 6 1 4 4 7 100 C 212 F 100 C 212 F 40 B hler Technologies GmbH BX460041 03 2021...

Page 41: ...GAS 222 20 Ex1 4 7 1 4 7 2 30 2500 DC 10 M 5 5 1 6 www buehler technologies com 7 www buehler technologies com 41 B hler Technologies GmbH BX460041 03 2021...

Page 42: ...GAS 222 20 Ex1 1 Appendix 1 1 Probe Terminal Diagram Probe heater Blowback solenoid valve Sample gas solenoid valve i i 42 B hler Technologies GmbH BX460041 03 2021...

Reviews: