AGF-VA-350, K-AGF-VA-350

• Observe the grade of filtration of the fine mesh filter and

if necessary install a pre-filtration to ensure larger solid
particles do not clog the filter prematurely.

• Observe the durability list in this document. Only use me-

dia compatible with the filter materials.

• Do not repair the filter. Damaged filter components must

be replaced.

• Do not paint, laminate or otherwise coat the filters.
• Observe the applicable constructor regulations, e.g. EN

60079-14, for installation and mounting.

• All metallic parts of the filter must be connected to an

earth potential. The resistance of a continuity test must
not exceed 1 MΩ (1*10

6

 Ω).

• The sealing materials, e.g. Teflon tape, must be conduct-

ive so all parts of the filter are earthed. If necessary, install
an earth bridge.

Operation and Control

DANGER

Risk of explosive gasses leaking and crossing zones

Check the filter is tight prior to operation.

Inadequately sealed filters can leak. Gas may leak or the am-
bient atmosphere be drawn in. Use a suitable method for the
leakage test, appropriate for the application.

4 Transport and storage

Only transport the product inside the original packaging or a
suitable alternative.

The equipment must be protected from moisture and heat
when not in use. They must be stored in a covered, dry and
dust-free room at a temperature between -20 °C to 50 °C (-4 °F
to 122 °F).

5 Installation and connection

5.1 Requirements to the installation site

The filter should be installed in a way so the filter element can
be replaced. If the filter protrudes from a contour, please note
this poses a risk of damage.

The spacing for the mounting holes (for M6 screws) is 25 mm.

The maximum permissible pressure is 270 bar abs. at temper-
atures up to 150 °C for models (K-) AGF-VA-350-T, or 200 bar
abs. at temperatures up to 240 °C for model AGF-VA-350-P.

5.2 Connecting a bypass

A 1/4” NPT connection thread for connecting a bypass is loc-
ated inside the filter cover.

• Screw in fitting.
• Perform a leak test with suitable means.

5.3 Connecting the gas lines

The connections must be made carefully and properly using
suitable fittings (gas connections right/left and bottom con-
densate drain/bypass: NPT 1/4”) and sealant.

When not using the condensate drain/bypass, attach a seal-
ing plug to the thread (sold separately - see appendix).

An arrow on the filter indicates the flow direction. The filter
can be screwed onto the mounting bracket to switch the inlet
and outlet side.

Perform a leak test with suitable means.

6 Operation and control

NOTICE

The device must not be operated beyond its specifications.

7 Service and Repair

Please refer to the original operator’s manual on the included
CD or online at www.buehler-technologies.com for a detailed
description of the unit including information on
troubleshooting and repair.

8 Maintenance

Always observe the applicable safety- and operating regula-
tions when performing any type of maintenance. Please refer
to the original operator’s manual on the included CD or online
at www.buehler-technologies.com for maintenance informa-
tion.

5

Bühler Technologies GmbH

BX410013 ◦ 06/2019

Summary of Contents for AGF-VA-350

Page 1: ...0 Dok No BX410013 Version 06 2019 Ref 998BX410013 Analysentechnik Kurzanleitung Filter deutsch 2 Brief Instructions Filter english 4 Notice de montage Filtre fran ais 6 Gu a r pida Filtro espa ol 8 ch...

Page 2: ...Giftige tzende Gase Das durch das Ger t geleitete Messgas kann beim Einatmen oder Ber hren gesundheitsgef hrdend sein a berpr fen Sie vor Inbetriebnahme des Ger ts die Dich tigkeit ihres Messsystems...

Page 3: ...ockenen und staubfreien Raum bei einer Temperatur von 20 C bis 50 C 4 F bis 122 F aufbewahrt werden 5 Aufbauen und Anschlie en 5 1 Anforderungen an den Aufstellort Der Filter sollte so eingebaut werde...

Page 4: ...ing it a Check tightness of the measuring system before putting it into operation b Take care that harmful gases are exhausted to a save place c Before maintenance turn off the gas supply and make sur...

Page 5: ...should be installed in a way so the filter element can be replaced If the filter protrudes from a contour please note this poses a risk of damage The spacing for the mounting holes for M6 screws is 2...

Page 6: ...tions d installation nationales en vigueur sont respect es DANGER Gaz toxiques ou irritants Le gaz de mesure transport par l appareil peut tre nocif pour la sant s il est inspir ou s il entre en conta...

Page 7: ...d exploitation doivent tre prot g s de l humidit et de la chaleur Ils doivent tre stock s dans une pi ce couverte s che et sans poussi re une temp rature comprise entre 20 C et 50 C 4 F bis 122 F 5 As...

Page 8: ...acionales de instalaci n PELIGRO Gases t xicos y corrosivos El gas de medici n conducido por el aparato puede resultar perjudicial para la salud al inhalarlo o al entrar en contacto con la piel a Ante...

Page 9: ...servar en un espacio a cubierto seco y li bre de polvo con una temperatura de entre 20 C a 50 C 5 Construcci n y conexi n 5 1 Requisitos del lugar de instalaci n El filtro debe instalarse de forma que...

Page 10: ...VA 350 K AGF VA 350 IIA IIB IIC 1 2 1 2 1 x 2 a b c 3 K AGF VA 350 AGF VA 350 1 2 IIA IIB IIC 2014 34 EU CE II 2 2G c IIC TX X 5 C Tamb 60 C 5 C 60 C 100 C AGF VA 350 K AGF VA 350 4 EN 60079 14 1 M 1...

Page 11: ...AGF VA 350 K AGF VA 350 5 2 NPT 1 4 5 3 NPT1 4 6 7 CD www buehler technologies com 8 CD www buehler technologies com 11 B hler Technologies GmbH BX410013 06 2019...

Page 12: ...stra e 29 40880 Ratingen Deutschland 49 0 21 02 49 89 0 49 0 21 02 49 89 20 1 1 AGF VA 350 K AGF VA 350 1 2 IIA IIB IIC 1 2 1 x 2 a b c d 3 K AGF VA 350 AGF VA 350 1 2 IIA IIB IIC 2014 34 CE Ex II 2 2...

Page 13: ...079 14 1 M 1 106 4 20 C 50 C 4 F 122 F 5 5 1 M6 25 270 150 C K AGF VA 350 T 200 240 C AGF VA 350 P 5 2 NPT 1 4 5 3 NPT 1 4 6 7 www buehler technologies com 8 www buehler technologies com 13 B hler Tec...

Reviews: