TC-Standard (+)
5.4 Description of menu functions
5.4.1 Display menu
Block temperature display
Display →
ch1
Depending on the device state, the temperature will be displayed as a constant, flashing, or alternat-
ing with a status message.
Ambient temperature display
Display →
ch2
The display is only available on devices with „Delta-T" option. Depending on the device state, the
temperature will be displayed as a constant, flashing, or alternating with a status message.
Operating hours /runtime display
Display →
oper
Displays the operating hours of the unit. The runtime cannot be reset and can be output in various
display formats. To view/exit the runtime, press the „Enter" key.
–
yy.mm
– display in years and months (default)
–
mth
– display in months
–
weeh
– display in weeks
–
days
– display in days
– One month corresponds to 30 days. Press the „F" key to switch between the display formats. The
display will then first show the selected format as short text, then the duration.
Error code display
Display →
err
In the event of errors/malfunctions not related to operation, the error code indicates possible causes
and solutions.
5.4.2 Main menu
Peltier Cooler TC-Standard (tc.St)
Display →
tc.st
This will take you to the cooler target temperature and the tolerance range setting (alarm
threshold).
Global setting (ToP Settings)
Display →
toP
This menu is used to configure the global cooler settings.
19
Bühler Technologies GmbH
BE440026 ◦ 12/2020