7 Maintenance and repairs
69
NIRFlex N-500
Operation Manual, Version G
MyNadiaFiles
Send this file to your customer service organisation for evaluation. To archive the file, copy it to a
different folder.
7 .4
Lamp module replacement
The primary lamp module needs to be replaced quickly after failure. This action can be performed by a
trained operator.
After failure of the primary lamp module, the following instructive message appears:
NOTE
When the primary lamp is broken the instrument automatically switches to the secondary lamp. In
this case make sure to replace the primary lamp as soon as possible as the instrument does not
work anymore in case of failure of the secondary lamp and this lamp has to be replaced by a Buchi
service engineer.