4 Description of function
42
NIRFlex N-500
Operation Manual, Version G
The LEDs at the handle of the fiber optic probe indicate the status of the system and the results of the
prediction in routine operation.
Applications without calibration (just measuring spectra, e.g. acquisition of calibration spectra)
-
green LED flashing
measuring
-
green LED on
measurement finished
red LED on
green LED on
message on monitor
Application with calibration (routine use for predictions)
-
green LED flashing
measuring
red LED on
green LED on
message on monitor
-
green LED on
result conforms
red LED on
-
result does not conform
Because the fiber optic probe is used at a certain distance from the spectrometer, the length of the
fiber optic cables between the measuring cell (spectrometer) and the probe tip is about 2 m (special
versions with longer fiber optic cables are available on request).
4
3
2
1
7
8
9
10
11
12
13
14
6
5
a
Protection cover
b
LEDs: OK, not OK
c
START button
d
Handle
e
Monofiber fiber optic 2 m
f
Electrical connections
g
Light barrier
h
Holder for fiber optic
i
Internal reference
j
Optic 3
k
Detector
l
Motor
m
Optic 1
n
Optic 2