7 Maintenance and repairs
72
NIRFlex N-500
Operation Manual, Version G
Workflow to reset the lifetime counter of the primary lamp module
To establish the standard working conditions of the NIRFlex N-500:
a
b
c
• Start the Operator software
• Switch from “secondary lamp”
a
to “primary lamp”
• Resetting the hour
b
counter will bring up a prompt for confirmation
• Confirm the message
c
to finally set the lamp lifetime counter to zero
➡
After this procedure, reboot the NIRFlex N-500 and restart the Operator software.