8 Troubleshooting
57
KjelFlex K-360
Operation Manual, Version D
To carry out an Operational Test proceed as follows:
• Set the system offline by pressing the corresponding button.
• Go to the part you want to test, e.g. Steam valve and switch it on by pressing the forward or the
backward button.
• If the valve works properly, a press sound is audible.
• Switch the valve off.
• Set the system online again.
NOTE
The button „Sensors“ enables to switch to the submenu Test Sensors.
8 .2 .2
Test sensors
This submenu enables to test the sensor functionality.
To test e.g. the tube shield sensor, proceed as follows:
• Open the tube shield. When the sensor is working properly “Open” will be indicated on the display.
• Close the tube shield. Now “Close” has to be indicated on the display.
NOTE
The button „Actors“ enables to switch to the submenu Operational Tests.
8 .2 .3
Test Printer
This submenu enables to test the functionality of the printer by e.g. printing a test page.
8 .2 .4
Operating Hours
This submenu indicates the operating hours of the steam generator, the number of distillations and the
date and time of the last reset.
8 .2 .5
Hardware Info
This submenu indicates some hardware information, e.g. part and serial numbers and the production
date of the instrument.
8 .2 .6
Service Test
This submenu enables the instrument manufacturers or service technicians to carry out a functionality
test of the instrument.