8 Troubleshooting
54
KjelFlex K-360
Operation Manual, Version D
8 Troubleshooting
This chapter helps to resume operation after a minor problem has occurred with the instrument. It lists
possible occurrences, their probable cause and suggests how to remedy the problem.
The troubleshooting table below lists possible malfunctions and errors of the instrument. The operator
is enabled to correct some of those problems or errors by him/herself. For this, appropriate corrective
measures are listed in the column “Corrective measure”.
The elimination of more complicated malfunctions or errors is usually performed by a Buchi technical
engineer who has access to the official service manuals. In this case, please refer to your local Buchi
customer service agent.
8 .1
Malfunctions and their remedy
Table 8-1: General malfunctions and their remedy
Error number / Error indication
Possible cause
Corrective measure
0
The current process was stopped.
Press OK, aspirate manually and start
again
1
Steam Valve (Y1 on) malfunction.
Contact the Buchi customer service
2
Aspiration Out Valve (Y2 on) malfunction.
Contact the Buchi customer service
3
Aspiration In Valve (Y3 on) malfunction.
Contact the Buchi customer service
4
Receiver Valve (Y4 on) malfunction.
Contact the Buchi customer service
5
Cooling Water Valve (Y5 on) malfunction.
Contact the Buchi customer service
8
Injection Valve (Y8 on) malfunction.
Contact the Buchi customer service
9
Water Valve (Y9 on) malfunction.
Contact the Buchi customer service
11
Steam Valve (Y1 off) malfunction.
Contact the Buchi customer service
12
Aspiration Out Valve (Y2 off) malfunction.
Contact the Buchi customer service
13
Aspiration In Valve (Y3 off) malfunction.
Contact the Buchi customer service
14
Receiver Valve (Y4 off) malfunction.
Contact the Buchi customer service
15
Cooling Water Valve (Y5 off) malfunction.
Contact the Buchi customer service
18
Injection Valve (Y8 off) malfunction.
Contact the Buchi customer service
19
Water Valve (Y9 off) malfunction.
Contact the Buchi customer service
20
H
2
O Pump (M1 on) malfunction.
Contact the Buchi customer service
21
NaOH Pump (M2 on) malfunction.
Contact the Buchi customer service
22
Reagent Pump (M3 on) malfunction.
Contact the Buchi customer service
23
H
3
BO
3
Pump (M4 on) malfunction.
Contact the Buchi customer service
25
H
2
O Pump (M1 off) malfunction.
Contact the Buchi customer service
26
NaOH Pump (M2 off) malfunction.
Contact the Buchi customer service
27
Reagent Pump (M3 off) malfunction.
Contact the Buchi customer service
28
H
3
BO
3
Pump (M4 off) malfunction.
Contact the Buchi customer service
30
Calibration out of range, check pump.
Check connection K-360/tank, check pump
33
The service door is open. If a process was
in progress, it may have been stopped.
Close the service door, start again
34
The splash protector door is open. If a
process was in progress, it may have been
stopped.
Close the splash protector door, start again
Summary of Contents for KjelFlex K-360
Page 1: ...KjelFlex K 360 Operation Manual 093176D en...
Page 71: ......