4 Description of function
21
KjelFlex K-360
Operation Manual, Version D
Fig. 4.3: Stop-cock
The stop-cock is located at the inside of the service door.
To carry out a Kjeldahl application the stop-cock must be in downwards position.
To carry out a non-Kjeldahl application the stop-cock must be in horizontal position.
!
CAUTION
Risk of burnings by hot machine parts and water steam.
• Let the system cool down before performing any service.
• Do not operate the instrument when enclosure is open or removed.
Notice
Wrong stop-cock setting will result in system overpressure. Tubing may come off.
• Set stop-cock to correct working-position.