6 | Operation
BÜCHI Labortechnik AG
20/30
Operation Manual HydrolEx H-506
u
Put 2 g Celite® 545 into the hydrolysis vessel.
u
Weigh-in sample.
u
Add 50 mL 4 M HCl and mix the sample with the
Celite® and the HCl.
u
Rinse the walls of the hydrolysis vessel with 50 mL, 4 M
HCl.
NOTE
Use quartz sand with a grain size of 0.3 - 0.9 mm.
4
In case the grain size is too small it will clog the frit of the glass sample tube.
4
In case the grain size is too large the Celite® will pass through the frit of the sam-
ple tube during filtration.
u
Apply approx. 20 g quartz sand equally in the glass
sample tube.
u
Add a layer of 2 g Celite® 545 in the glass sample tube.