5 | Installation
BÜCHI Labortechnik AG
16/30
Operation Manual HydrolEx H-506
NOTICE
Risk of property damage and diminished performance due to use of un-
suitable power cables.
The power supply cables supplied with the product by BUCHI precisely match the re-
quirements of the device. If other power cables that do not meet those requirements
are used, the device may be damaged and/or its performance diminished.
u
Use only the power supply cables supplied with the product or ordered separately
from BUCHI.
u
If using any other power supply cables, make sure that they match the specifica-
tions on the type plate.
u
Make sure that all connected devices are grounded.
u
Make sure that the power plug is freely accessible at all times.
u
Insert the power cable into the connection labeled
Power IN
on the back of the in-
strument.
u
Insert the power plug into the power socket.
5.4
Securing against earthquakes
NOTE
4
Use two eye screws M4 x 10.
4
Screw-in depth 10 mm.
u
Attach two eye screws to the instrument.
u
Attach the instrument to a fixed point using a
strong cord or a wire.
5.5
Assembling the suction tube
NOTE
Malfunction because of incorrect assembled suction tube.
Make sure:
4
That the stopcocks are freely accessible.
4
That the openings are facing upright.
u
Switch the
On/Off
master switch to Off.