36
BÜCHI
Fat Determination
B-820/B-821
5.2.3 System configuration
Position the cursor on <Setup>.
The cursor is positioned on <Gas flow setup>. Position the
cursor on <System configuration>.
Balance
The cursor is positioned on < Balance>.Use the OPTION key
to select parameters for the following balances: Sartorius (MC1
types as well as balances with interface MP8-4), Mettler AE,
Mettler AT, A & D (FR, HA series) or none.
Frequency
The cursor is positioned on <Frequency>. Use the OPTION
key to select the network frequency (50 Hz or 60 Hz).
Column
The cursor is positioned on <Column>. Use the OPTION key
to select between the fast "total fat" determination with the
short column and the "fatty acid" determination with the long
column.
The cursor is positioned on <
Time/Date
>.
•
The date can be entered
(German: DD.MM.YY; English: MM.DD.YY).
•
The correct time can be entered (hours, minutes).
Position the cursor on <
PC interface
>.
Gas flow setup
User configuration
Printer configuration
System configuration
Autosampler
Main menu
Preheating
p H2
55kPa
Sample weight/Mode
p air
50kPa
Determination
T oven
130°C
Data manager
T inj.
220°C
Setup
T FID
260°C
Balance:
Mettler AE
Frequency:
50Hz
Column:
Total Fat
Time/Date
PC-Interface
Main menu
5 Operation