background image

48

BÜCHI 

Fat Determination

 B-820/B-821

5 Operation

Sample no:          5

p  H2

5 5 k P a

Next number

p air

5 0 k P a

T oven

8 9 ° C

T inj.

2 2 0 ° C

Main menu

T FID

2 6 0 ° C

Wait until ready

 spec.corr.

Name:    cream cheese

Sample  no:                      5

p  H2

5 5 k P a

Next number

p air

5 0 k P a

T oven

9 0 ° C

T inj.

2 2 0 ° C

Main menu

T FID

2 6 0 ° C

Ready injection   2

  spec.corr.

Name:      cream cheese

Figure 21b:
Fat determination report (short column)
without printout of the vials; internal
printout

Figure 21c:
Factor determination report (long column)
without printout of the vials; internal
printout

Determination with manual syringe

Sample No.

Entry of the sample number (from the sample weight table)
that should be determined as the next sample. is Confirm this
with “ENTER”; the cursor moves to the next row. The
corresponding sample identification (Name), the injection
number (here the second injection) and the sample-specific
determination mode (“spec.corr.”) will be shown in the display.

If the set temperatures and pressures have not been reached,
or as long as “Preheating” is running, the message “

Wait

until ready

” will be shown on the display. In this case, no

LED will be illuminated, or the “Running” LED will be on.

Procedure

If the set values are reached, the “READY” LED lights up,
and the message “

ready injection

” appears in the display

(in this case, the second injection). After carefully rinsing the
syringe with n-butanol, 2 ml of extract (upper phase) are now
injected into the injector port using the microliter syringe.
The temperature program starts automatically after the
injection. The corresponding temperature program runs
automatically in the background according to the determination
mode (see Chapter 3, Table 7). At the end of the determination,
the venting flap opens; the oven area will be conditioned in
order to prepare the system for the next determination. The
evaluation takes place automatically, and will be printed out
depending on the printer configuration (see Chapter 5.2.2).
The next manual injection can take place as soon as “Ready
for Injection” appears in the display (the “READY” LED lights
up).

Summary of Contents for B-820

Page 1: ...bols 7 2 2 Requirements for the customer 7 2 3 Proper use 7 2 4 Improper use 8 2 5 General hazards 8 2 6 Safety measures 9 2 7 Handling of gas cylinders in the laboratory 10 2 8 Modifications 10 2 9 R...

Page 2: ...on safety which must be observed if the Fat Determi nation Unit is to be operated safely Due to continuous improvement and updating technical spe cifications may be altered without prior notice No par...

Page 3: ...924 37626 Automatic basic system incl individual Version B 815 B 820 B 821 Total fat order no fatty acids order no 230V 230V 230V 230V 36923 37631 200V 120V 200V 230V 37108 37630 120V 120V 230V 230V 3...

Page 4: ...umn Fatty Acids 37620 Calibration Set Fatty Acids 37622 EPROM Set manual or automatic Accessories B 820 Manual 1 Soap bubble meter compl 34878 1 Separation column Total Fat 34916 1 Supply line H2 3613...

Page 5: ...2 Fork spanners w a flats 8 16522 1 Fork spanner w a flats 17 x 19 19831 1 T piece D1 8 36845 1 Jumper cord W VDE 30983 1 Power cable PNE 2 5 m available in the following versions Type CH 10010 Type G...

Page 6: ...syringe 37067 1 Tubing PUR 1 8 L 3000 1 Foam rubber of neoprene length 100mm Table 4 Accessories Autosampler B 821 Compare with spare parts list of the original operating instructions for the 8200 CX...

Page 7: ...zards that can arise when using the unit Personnel without such training or persons currently under going a course of training should be trained thoroughly This Instruction Manual should be the basis...

Page 8: ...f gas cylinders inadequate installation of connections to gas cylinders Apart from authorized maintenance personnel no one is allowed to remove any safety devices and covers Touching live parts may re...

Page 9: ...anger cannot be ensured any more if the unit shows visible damages the unit does not work any more after long term storage under inadequate circumstances after having been exposed to considerable stra...

Page 10: ...re Close cylinder valves immediately Cool any overheated cylinder with plenty of water 8 Only keep as many replacement cylinders in the laboratory as are necessary for continuous operation 9 Open cyli...

Page 11: ...atrices using a gas chromatographic technique It is designed to function with the B chi Extraction Unit B 815 The temperature program and the evaluation function are preprogrammed and have an extremel...

Page 12: ...tronic Output Main principle 1 Switch on the unit the switch is located on the back of the unit Perform the input steps as described in Section 5 Operation 2 Using a syringe extract some solution from...

Page 13: ...ration FA calib Determines the retention period for six saturated fatty acids calibration set Fatty Acids Fatty acid factor FA factor Determines the factor that will be necessary for the determination...

Page 14: ...ection to Autosampler B 821 to adjust your own country s voltage Turntable Pneumatics module Power on off switch Power entry module Carousel support Carousel Injection tower module 3 2 2 Carousel supp...

Page 15: ...ate the unit are not included in delivery and cannot be ordered from the manu facturer Hydrogen percentage purity 5 0 gas cylinder or central supply Synthetic air gas cylinder or central supply 2 pres...

Page 16: ...unit If a balance or printer is connected the proper balance or printer type must be entered in the initialization menu under system configuration or printer configuration see Section 5 2 3 System co...

Page 17: ...anguage LQ Epson Emula tion are supported Epson Canon HP DeskJet HP LaserJet The following printers have been tested at B chi Epson Stylus Color 440 Canon Bubble Jet BJC 600 BJ 200 EX DIP switch setti...

Page 18: ...s arms Only fill when the unit is turned off and is at room temperature Filling under any other circumstances may lead to system contamination 4 6 Gas connections Gases used Gas function Gas type Sett...

Page 19: ...the B 820 See Figure 10 Gas connection layout Page 19 Ensure the correct connection of the gas cylinder to the unit The following are included in the B chi delivery H2 connection inc copper capillary...

Page 20: ...tor on the pressure reducing valve should remain constant for approx 10 minutes The cylinder valves can be reopened once this check is completed The primary pressure the tightness of the supply lines...

Page 21: ...eration The procedure is the same for the Fatty Acid column Figure 12 Installing the colums The column should be heated only when gas H2 is flowing Prior to operation always ensure that the gas supply...

Page 22: ...red it can also be carried out manually by pressing key combination SHIFT i From the top of the FID look for red heating coil If the heating coil is defective use a lighter to ignite the FID flame 4 1...

Page 23: ...he B 820 unit In order to mount the Autosampler on the B 820 unit proceed as follows 1 Place the Autosampler on the mounting plate just mounted Push the Autosampler carefully to cover the position hol...

Page 24: ...bs down with the other hand Check that the mounting is fully closed before pushing the side locking tabs down It is closed when the red signal tab is not visible 4 Connect the solvent line laterally t...

Page 25: ...In order to align the syringe needle with the injection head of the B 820 unit proceed as follows 1 Check that the Autosampler is OFF 0 Press down the latch refer to Figure 16 Carousel support of the...

Page 26: ...n flap Affix it to the top of the B 820 4 11 11 Installing the solvent reservoir bracket The bracket holds two solvent bottles of 180 ml each and is secured on the lower left side of the B 820 unit by...

Page 27: ...not routed in front of the air flap of the oven in the back of the unit 2 The lines are routed along the bottom of the B 821 unit up to the solvent reservoirs 3 Fill a bottle of n pure butanol or rin...

Page 28: ...n be operated by turning it on once again The operating panel shows Display Lamp lights up when the unit is ready for an injection or a gas flow measurement Lamp lights up when the determination or pr...

Page 29: ...command Allows selection of a variety of options e g balances print system frequency etc In the sample weight table pressing the OPTION key copies a field into the next row These keys enable input of...

Page 30: ...30 B CHI Fat Determination B 820 B 821 Use the keyboard to enter sample identification and report titles and press this key to cancel The key has the same function as this key ENTER 5 Operation...

Page 31: ...ing the unit The following flow chart shows the operating structure This is followed by a systematic description of all the options provided by this unit The unit contains a help program which can be...

Page 32: ...FID 260 C Determination with autosampler Determination without autosampler Sample no 5 p H2 55kPa Next number p air p air 50kPa T oven 130 C T inj 220 C Main menu T FID 260 C Ready injection 2 gen cor...

Page 33: ...onfiguration System configuration Autosampler Main menu Gas flow setup for H2 short column User configuration Printer configuration System configuration Baudrate 9600 Data Bits 8 Stop Bits 1 Parity no...

Page 34: ...rman French Italian or Spanish Position cursor on User The user name can be entered Position cursor on Unit The following parameters may be chosen g 100g g kg mg kg ppm Only the g 100g unit is activat...

Page 35: ...will subsequently be printed on each report Position cursor on Online intern Use the OPTION key to activate yes or deactivate no the internal printer to print the result after determination Position c...

Page 36: ...50 Hz or 60 Hz Column The cursor is positioned on Column Use the OPTION key to select between the fast total fat determination with the short column and the fatty acid determination with the long col...

Page 37: ...ss the OPTION key to select the number of bits for the transmission Position cursor on Stop bits Press the OPTION key to select 1 or 2 Press the OPTION key to position parity even odd or none Position...

Page 38: ...es are reached the heating elements begin to heat up The cursor is positioned on Preheating The actual temperature and pressure values are displayed on the right pH2 Actual pressure for hydrogen pair...

Page 39: ...procedure press the ESCAPE key 5 2 6 Gas flow rate setup If the unit has been turned off for longer than 30 minutes the unit must be preheated for 30 minutes before the gas flow rate setup procedure...

Page 40: ...ressure must be adjusted accordingly Position the cursor on H2 pressure and enter new value Once the ready LED lights up the flow rate can be remeasured This procedure should be repeated until the set...

Page 41: ...ould be 2 7 0 1 sec flow rate 20 ml 2 7 0 1 sec 7 If the stopped time deviates from the set time the air pressure value must be adjusted accordingly position the cursor on air pressure and enter new v...

Page 42: ...watch counting in 1 100 th of a second interval 10 ml Marker 3 20 ml Marker 2 Marker 1 5 Operation Re check the gas flow rate and the supply lines for leakages see Section 4 7 Checking supply lines f...

Page 43: ...nd the internal standards for the connected balance type Depending on the mode selected the cursor will automatically move to the next available field Preheating p H2 55kPa Sample weight Mode p air 50...

Page 44: ...e OPTION key Within the column you can move between the fields of each sample number with SELECT You can move to the first sample number No 1 with the key combination SHIFT SELECT upward arrow and to...

Page 45: ...ace the cursor on Main Menu with SELECT and confirm with ENTER Or press ESCAPE twice after completing the entries to move directly to the Main Menu All the inputs made in Factors must be updated by th...

Page 46: ...equivalent to a STAND BY mode energy savingmode Ifthesystemfindsitselfin STAND BY after the last determination the display will indicate The determination finished correctly The determination finishe...

Page 47: ...r sample will be entered under Autosampler in the Setup menu see Chapter 5 2 4 A maximum of 9 injections per sample vial are provided for the Total fat option and a maximum of 6 injections per sample...

Page 48: ...he set temperatures and pressures have not been reached or as long as Preheating is running the message Wait until ready will be shown on the display In this case no LED will be illuminated or the Run...

Page 49: ...le Entry of the first sample number of wich the report shall begin Confirm entry with ENTER Last sample Entry of the last sample number of the report Confirm entry with ENTER Start output Start the ou...

Page 50: ...s and the contents for FA TG Fatty Acid Trigylceride of each injection will be printed out see Fig 21e In all other modes the sample and user data and the results of each injection will be printed out...

Page 51: ...ion i 1 2 m with m 1 2 75 Fatty acids option i 1 2 m with m 1 2 48 n Number of injections carried out per sample Total fat option n 1 2 9 Fatty acids option n 1 2 6 m Number of samples that should be...

Page 52: ...desired type of output unit can be selected with OPTION Possible selections OPTION for the device Device Display Display of the data on the display Display is the default setting internal printer pri...

Page 53: ...number e g 5 1 6 1 the number of the injection e g 5 1 6 1 and the sample name The cursor can be moved to the desired sample name measurement using the SELECT key In the Elim column a measurement can...

Page 54: ...efined interval first to last sample number After placing the cursor on Clear data confirm the function with ENTER Sample data will be actively cleared and the message Data has been cleared appears Pl...

Page 55: ...100 injections These can be ordered from B chi When a column is replaced a new septum must be installed Only replace the septum when the unit is turned off The unit must be in a room temperature pres...

Page 56: ...er No 036678 jet sits inside the FID s base body 7 Clean jet with fine wire and clean the collector if necessary 8 Reinstall FID ensure that all sealings are reinstalled 6 4 Replacing the ignition coi...

Page 57: ...s at room temperature Filling under any other circumstances may lead to system contamination 6 6 Installing a new separation column Loosen the threaded column joint and carefully remove the old column...

Page 58: ...uthorized service personnel i e skilled technicians who are familiar with the operating instructions and current safety regulations and accident prevention B chi service centers have instrument specif...

Page 59: ...or discoloured the syringe must be rinsed once again using a suitable solvent Needle replacement To replace the needle proceed as follows Loosen the knurled nut Remove the old needle from the syringe...

Page 60: ...k of the unit 7 2 Storage Transport The units should be stored and transported in their original packaging and in clean condition Ensure that the separation column has been removed from the oven of th...

Page 61: ...arts found in this manual is also prohibited Replacement parts detection unit B 820 Order no Paper roll 26341 Printer ribbon 26400 FID ignition coil 27275 Syringe 10 l 34258 Needle D5 0 18 36654 Gas s...

Page 62: ...W Current consumption max 9A Ambient temperature 5 40 C Table 10 Technical data B 820 B 821 Unit dimensions 155 x 495 x 400 mm width x height x depth Packaging dimensions 630 x 460 x 680 mm width x he...

Page 63: ...ernal printer does not function Defective communication between the printer and the electronics Please contact your local B chi Service office or B chi Labortechnik AG see Chapter 6 10 Error 09 The co...

Page 64: ...e Autosampler B 821 is faulty make one or more injections by hand If the problem remains check the Autosampler While it is frequently assumed that the sampler is defective the problem may rather be ca...

Page 65: ...ference in which the user will be required to correct the interference at his own expense Fran ais Cet appareil a t test et s est av r conforme aux limites pr vues pour les appareils num riques de cla...

Page 66: ...c compatibility Generic immunity standard Residential commercial light industry EN 60555 2 1987 IEC 555 2 Disturbances in supply systems caused by hosehold Rotavapors and similar electrical equipment...

Reviews: