Inserting the Flow Cell
H12214_1_003
19
6
Inserting the Flow Cell
Before inserting the Flow Cell make sure the CryoFIT Accessory has been assembled
correctly as described in the section
Assembling the CryoFIT Accessory [
.
• Switch off the BCU05 to prevent freezing.
• Close the valve to disable the spinning flow. Move the slider to the closed position.
Figure 6.1: Valve Parts
• Activate the sample lift.
CAUTION! Make sure the sample changer is not used during the flow application.
Figure 6.2: Activating the Sample List
Note
: The insert guide is needed to insert the capillary properly into the probe. The insert
guide looks similar to a classical sample spinner, nevertheless, never try to spin it!