Installation
16
H12214_1_003
5.2
Installation with a Cold CryoProbe
Make sure firmware version
crcoah
or higher is installed before installation.
• Dismount the mechanical Tuning & Matching adapter or the ATM.
The error message:
(152) Temp. controller #3 sensor missing
will appear on the
display.
NOTICE! Proceed to the next steps without delay!
• Dismount the VT – gas tube from the CryoProbe.
• Plug in the T – piece for the VT gas and capillary on the CryoProbe.
• Mount the ATM unit or the mechanical Tuning & Matching adapter.
The CryoProbe is now ready again for use with NMR sample tubes. To use the flow cell,
follow the instructions in the following chapters of this manual.
5.3
Assembling the CryoFIT Accessory
Before using the CryoFIT Accessory the valve, bracket and holder unit must be mounted.
• Mount of the valve used to disable the spinning flow on the hose leading to the Shim
Upper Part:
Figure 5.2: Valve Mounting