5.3
Checklist of Installation Requirements
Table 5.2:
Installation requirement checklist
One cylinder of Nitrogen gas.
One cylinder of Helium gas.
Quantities of liquid Helium and Nitrogen in transport dewars.
Three power sockets (230 V / 16 A single phase or 208 V / 20 A single
phase or 110 V / 20 A). The power outlets will be used to run a vacuum
pump, a heat gun and a power supply unit. These power outlets must be
available in addition to the power source used to run the spectrometer.
Step ladder (non-magnetic e.g. aluminum or wood).
Where possible the customer should provide a heat gun or powerful
handheld air dryer (min. 800 W), a roughing pump (10
-2
mbar) and a pair
of insulated gloves. If those items can not be provided by the customer,
then Bruker can provide them.