Lab Requirements
4.1
Room Layout
The Fourier system can be placed in to a small room area but certain requirements have
to be met. There should be a clear path, free of obstacles, to bring in containers to
service the magnet with liquid Nitrogen and Helium. The console can be placed close to
the magnet but it is recommended that the work station should be placed a minimum
distance away from the magnet. As a room layout example, the Fourier system with the
optional SampleXpress Lite is shown in Figure 4.1 below.
Figure 4.1:
Room layout example of an installed Fourier system
4.2
Ceiling Height
A
minimum ceiling height of 2.74 m (9 feet)
is required for the clearance needed
above the magnet for assembly, energizing and filling liquid Helium.