24
B127527_00_03
Lab Requirements
4.9.1
Checklist for Utility Requirements
Table 4.5:
Utility requirement checklist
Provision must be made for any additional electrical power requirements
for extra equipment that will be installed.
Additional electrical outlets are required for turbo-pumps and magnet
power supply during installation and service.
A voltage stabilizer must be available if the line voltage fluctuation
exceeds -5% to +10%.
An UPS should be considered if power outages are frequent.
The power supply to the spectrometer should be clean.
A minimum of one compressed gas line with one regulated output must
be available.
Provision must be made for gas requirements of any accessories used.