CHAPTER 3
Start-up and Testing
39
Fig.3.2
One Controller module
and one Input module
Fig.3.3
Multiple modules in a
multi-frame setup
6) Click
Start
.
7) If you have not checked
I
nternal cabling
, connect cables as the program indicates.
This will test your system all the way from the input/output connections to the Win-
dows
®
application.
The program will advise you when the specific module has passed or failed an analog
self-test (Fig.3.4). View the test Log to read any failure messages and, if desired, select
File
,
Save
or
Save as
to save the Log in rich text format. Click
Help
for more information.
When all present modules have passed the test, you should have a working system
ready for measuring.