4. Preventive Maintenance
User Manual
Cleaning the Cold Probe
6. Inspect the condenser face about weekly for the first month then monthly thereafter to determine
the cleaning frequency the ventilated panel needs cleaning.
Cleaning the Cold Probe
This procedure describes how to clean the cold probe.
WARNING
THERMAL HAZARD (cold)
Do not come in contact with the probe or baffles while they are
cold. Frost bite or other contact could cause personal injury.
Wait at least 90 minutes after the P-102 is turned off before
allowing the probe or baffles to touch your body.
Tools and Materials
Isopropyl alcohol
Clean wipe or lintless cloth
1. Turn off the P-102 and unplug the unit.
2. Perform Lockout/Tagout according to the facilities procedure.
3. Wait for 90 minutes for the cold probe to return to room temperature.
4. Wipe the cold probe with isopropyl alcohol wipes until it is clean.
Accelerating Innovation
4-28