4. Preventive Maintenance
P-102 Maintenance
User Manual
4. Preventive Maintenance
P-102 Maintenance
Follow the preventive maintenance procedures and schedule provided in this section. Performing
preventive maintenance extends the operating life of the product. Adjust the schedule you need
depending on where the Polycold Cryocooler is used.
WARNING
Mechanical hazards, electrical hazards, and shock hazards
exist on the P-102. The procedures in this chapter should only
be performed by qualified persons. Read and understand the
Safety section in Chapter 1 before performing any procedure.
Parts
Polycold can provide all parts required for Preventive Maintenance. For a list of these parts, contact
Brooks Automation Technical Support. Refer to
Chapter 4: Parts and Service
for contact information.
Preventive Maintenance Schedule and Procedures
Always perform Lockout/Tagout according to the facility procedure before removing any covers or
handling the cold probe or baffle. Wait 90 minutes after Lockout/Tagout is performed before opening
covers or handling the cold probe or baffle.
WARNING
The fan has no obstruction sensors and can cause serious
personal injury.
Do not operate the Product without the protective covers in
place.
4-25