Standard Response and Current I/O PC Board Adjustment #2
(specification guaranteed)
Note: Adjustment of the anticipate potentiometer to obtain a flow rate
performance to be within 2% of flow rate commanded in less than three
seconds (six seconds for Current I/O Version) after command change
requires the use of a
fast response flowmeter
(500 millisecond response to
be within 0.2% of final value or better) in series with the Model 5850EM
and a storage oscilloscope or recorder.
a. Make a step change in command to the controller from zero to 100% of
full scale flow and record the output signal of the
fast response flowme-
ter
.
b. If this signal shows more than 4% overshoot, adjust the anticipate
potentiometer one-half to one turn counterclockwise. If the signal does
not show overshoot, but is not within 2% full scale of final value after
three seconds, adjust the anticipate potentiometer one-half to one turn
clockwise. Set the command potentiometer for zero percent of flow.
c. Repeat Steps a and b until the
fast response flowmeter
output signal
meets the specified response requirements.