Brookfield Engineering Labs., Inc.
Page 9
Manual No. M02-33E209
Table III-3: Accuracy for Low Torque CAP 2000+
Cone
≤00 RPM
70 RPM
900 RPM
900 RPM
10-100% FSR
10-100% FSR
≤
50% FSR
>
50% FSR
01
±2.0%
±2.0%
±2.0%
±2.0%
02
±2.0%
±2.0%
±2.0%
±2.0%
03
±2.0%
±2.0%
±2.0%
±2.0%
04
±2.0%
±2.0%
±2.0%
±2.0%
05
±2.0%
±2.0%
±2.0%
±2.0%
06
±2.0%
±2.0%
±2.0%
±2.0%
07
±2.0%
*
*
*
08
±2.0%
±2.0%
±2.0%
±2.0%
09
±2.0%
±2.0%
±2.0%
±2.0%
10
±2.0%
±2.0%
±2.0%
±2.0%
* Not recommended for use at these RPM
Table III-4: Accuracy for High Torque CAP 2000+
Cone
≤00 RPM
70 RPM
900 RPM
900 RPM
10-100% FSR
10-100% FSR
≤
50% FSR
>
50% FSR
01
±2.0%
±2.0%
±2.0%
±4.0%
02
±2.0%
±2.0%
±2.0%
±4.0%
03
±2.0%
±2.0%
±2.0%
±4.0%
04
±2.0%
±3.0%
±3.0%
±6.0%
05
±2.0%
±4.0%
±4.0%
±8.0%
06
±2.0%
±5.0%
±5.0%
±10.0%
07
±2.0%
*
*
*
08
±2.0%
*
*
*
09
±2.0%
*
*
*
10
±2.0%
±2.0%
±2.0%
±2.0%
* Not recommended for use at these RPM
The specification of temperature accuracy on CAP Viscometers is stated below:
LOW TEMP UNITS
:
• In ambient conditions of 15°C to 30°C, accuracy is +/- 0.5°C when the temperature set point
on the viscometer is 5°C to 75°C.
• In ambient conditions outside of 15°C to 30°C, accuracy is +/- 1.0°C when the temperature
set point on the viscometer is 5°C to 75°C.
HIGH TEMP UNITS
:
• In ambient conditions of 15°C to 30°C, accuracy is +/- 0.5°C when the temperature set point
on the viscometer is 50°C to ≤100°C.