In FlowPlot, the valve power parameters are set on the
Controller
tab of the
Instrument Settings
:
·
Valve maximum
sets the power needed to fully open the
valve
·
Valve open
sets the power needed to keep the valve open
(hold signal)
Optimum values might differ, depending on the specifics of
the valve.
If the use of the valve power parameters seems somewhat counterintuitive, note that usage here is specific to shut-off
valves. For control valves, the purpose of the
Open
and
Max
parameters is more straightforward.
The valve power settings can be done with the following digital parameters:
Name
Process/
Parameter
Modbus
register no
Value
Purpose/Remarks
Valve Maximum (231)
114/25
62153…
62154
as required set minimum power for opening shut-off valve
Valve Open (190)
114/24
62145…
62146
as required set minimum power to keep the valve open
5.4
Optimizing hints
Problem
Possible cause
Action
Batch size overshoot not reduced
after more than 3 batches
Dosing time too short
·
Decrease inlet pressure
·
Decrease
Counter control gain too high
Decrease
Dosing time too long
Flow rate too low
·
Increase inlet pressure
·
Increase
Counter control gain too low
Increase
Controller too slow
Optimize
Pump speed adjustment too slow
Increase controller speed (when using a pump,
controller speed values up to 10 can be used)
New batch cannot be started
New batch trigger(s) not defined
Enable new batch trigger(s) (see
Flow continues after configured
batch size is reached
New setpoint on limit
> 0%
Set
New setpoint on limit
to 0% (see
Counter value still changes after
limit is reached
Measuring signal disturbed by
electrical noise
Increase minimum capacity (see
Zero point drifted
Adjust zero point (see
Gas inclusion in tubing
Flush instrument and tubing at relatively high
flow rate before starting batch control