Bronkhorst®
Instruction Manual mini CORI-FLOW ML120
9.17.097G
30
Available functions during powering-up
·
In order to access this sequence, press and hold the switch while powering-up the instrument.
·
LED indications in this sequence are flashing (0.2 sec on, 0.2 sec off).
•
(green)
•
(red)
Time
Function
•
•
0…4 sec
No action
•
•
4…8 sec
Restore factory settings
(except communication settings)
•
•
8…12 sec FLOW-BUS:
Auto install to bus
; let the instrument obtain a free node address from
the FLOW-BUS system
•
•
12…16 sec
Activate configuration mode
:
·
The baud rate and bus type for the 9-pin D-sub side connector are set to 38k4 and
RS232 FLOW-BUS (ProPar)
·
In configuration mode, the green LED blinks 2 seconds on and 0.1 second off
·
Configuration mode remains active after powering-down and can be deactivated
by selecting this function again at the next start-up