Bronkhorst®
Instruction Manual mini CORI-FLOW ML120
9.17.097G
54
5.3.4
Disabling filters
Dosing speed can be increased by disabling output and
sensor filters. In FlowPlot, filter settings are located on the
Basic
tab of the
Instrument Settings
function. To disable
filtering, make sure to set the values as shown in the image
to the right.
Disabling filters, however, not only makes the instrument
faster, it also makes it more sensitive to vibrations and
electrical noise.
Minimizing sensitivity
The instrument should always be mounted in a vibration free environment (if possible), on the supplied mass block and
dampers. Sensitivity to electrical noise can be eliminated by increasing the minimum capacity of the instrument to a value
just above the noise level.
The minimum capacity cannot be entered in FlowPlot; instead, use the parameter test function of FlowDDE. To read and
write parameters with FlowDDE, proceed as follows:
1. In FlowDDE, open
FLOW-BUS > Test FLOW-BUS and DDE
2. From the
Parameter
list, select the required parameter
3. In the
Write value
column, enter the required parameter
4. Click the
Write
button to store the value
5. Click the
Read
button to check if the write action has succeeded
The value of
Capacity Minimum (247)
can only be changed after enabling edit mode for secured parameters with parameter
Init Reset (7)
. To set the minimum capacity, enter the required information as shown in the image below (the value of
Capacity Minimum
is illustrative, use the value you need).