Manual or Automatic Grand Total Reset Function (
GtrST
)
For manual reset, select
T RESET
Gtrst
man
and then select whether manual reset will be
enabled (
Enabl
) or disabled (
dsabl
) using the
Up
arrow key. Press the
Enter
button to accept. Disabling
reset will avoid inadvertent resets of the total via the front reset button.
For automatic reset, select
T RESET
Gtrst
Auto
T DELAY
and enter reset delay time in
seconds. Once the grand alarm output grand total set point is reached, the meter waits for a programmed
amount of time (
T DELAY
) and then resets the grand total to zero.
Press the
Enter
button, at any time, to accept a setting; otherwise press the
Menu
button to exit without
saving changes.
Grand Total Alarm Time
The
T DELAY
parameter is used by the NPN open collector outputs when they are programmed as
grand total alarms. If
grand total reset
(
Gtrst
) is programmed to
Auto
,the
time delay (
T DELAY
) is the
length of the associated Out 1 or Out 2 grand total alarm prior to the grand total being reset to 0.
For information on programming the NPN open collector pulse outputs as grand total alarms, see Alarm
Output programming on page 48.
Press the
Enter
button, at any time, to accept a setting; otherwise press the
Menu
button to exit without
saving changes.
60
Summary of Contents for BERT-E
Page 1: ...Model BERT E Installa on and Opera on Manual BI BERT Revision 01 Engineered for Accuracy...
Page 28: ...26...
Page 38: ...36...
Page 51: ...49...
Page 68: ...66...
Page 70: ...68...
Page 78: ...76...
Page 84: ...Mounting Dimensions All units inches mm Figure 14 Enclosure Dimensions Front View 82...
Page 90: ...This page left blank intentionally 88...
Page 91: ...Notes 89...
Page 92: ...Notes 90...
Page 93: ......