Total Reset (
T Reset
)
This menu is used to select the ways the total and grand total may be reset.
Manual or Automatic Total Reset Function (
t
rST
)
For manual reset, select
T RESET
t rst
man
and then select whether manual reset will be
enabled (
Enabl
) or disabled (
dsabl
) using the
Up
arrow key. Press the
Enter
button to accept. Disabling
reset will avoid inadvertent resets of the total via the front reset button or external reset contact.
For automatic reset, select
T RESET
t rst
Auto
T DELAY
and enter reset delay time in
seconds. Once the output alarm total set point is reached, the meter waits for a programmed amount of
time (
T DELAY
) and then resets the total to zero.
Press the
Enter
button, at any time, to accept a setting; otherwise press the
Menu
button to exit without
saving changes.
Total Alarm Time
The
T DELAY
parameter is used by the NPN open collector outputs when they are programmed as total
alarms. If
total reset
(
t rst
) is programmed to
Auto
the
time delay (
T DELAY
) is the length of the
associated Out 1 or Out 2 total alarm prior to the total being reset to 0.
For information on programming the NPN open collector pulse outputs as total alarms, see Alarm Output
programming on page 48.
59
Summary of Contents for BERT-E
Page 1: ...Model BERT E Installa on and Opera on Manual BI BERT Revision 01 Engineered for Accuracy...
Page 28: ...26...
Page 38: ...36...
Page 51: ...49...
Page 68: ...66...
Page 70: ...68...
Page 78: ...76...
Page 84: ...Mounting Dimensions All units inches mm Figure 14 Enclosure Dimensions Front View 82...
Page 90: ...This page left blank intentionally 88...
Page 91: ...Notes 89...
Page 92: ...Notes 90...
Page 93: ......